Proc Inst Mech Eng H ; 224(6): 751-63, 2010.
Article in English | MEDLINE | ID: mdl-20608492

ABSTRACT

This paper presents a novel wheeled probe for the purpose of aiding a surgeon in soft tissue abnormality identification during minimally invasive surgery (MIS), compensating the loss of haptic feedback commonly associated with MIS. Initially, a prototype for validating the concept was developed. The wheeled probe consists of an indentation depth sensor employing an optic fibre sensing scheme and a force/torque sensor. The two sensors work in unison, allowing the wheeled probe to measure the tool-tissue interaction force and the rolling indentation depth concurrently. The indentation depth sensor was developed and initially tested on a homogenous silicone phantom representing a good model for a soft tissue organ; the results show that the sensor can accurately measure the indentation depths occurring while performing rolling indentation, and has good repeatability. To validate the ability of the wheeled probe to identify abnormalities located in the tissue, the device was tested on a silicone phantom containing embedded hard nodules. The experimental data demonstrate that recording the tissue reaction force as well as rolling indentation depth signals during rolling indentation, the wheeled probe can rapidly identify the distribution of tissue stiffness and cause the embedded hard nodules to be accurately located.


Subject(s)
Elasticity Imaging Techniques/instrumentation , Minimally Invasive Surgical Procedures/instrumentation , Palpation/instrumentation , Surgery, Computer-Assisted/trends , Transducers , User-Computer Interface , Equipment Design , Equipment Failure Analysis , Reproducibility of Results , Sensitivity and Specificity , Touch
SELECTION OF CITATIONS
SEARCH DETAIL
...