Guang Pu Xue Yu Guang Pu Fen Xi ; 25(8): 1355-7, 2005 Aug.
Article in Chinese | MEDLINE | ID: mdl-16329521

ABSTRACT

A method is reported for the determination of trace bismuth in traditional chinese medicine by hydride generation atomic fluorescence spectrometry. The effect of different means of digestion, the medium amounts of acid and reducing agent on the determination of Bi is investigated. In the given conditions, the linear range of determination is 0.1-200 microg x L(-1), and the detection limit is 0.0946 microg x L(-1). The instrumental relative standard deviation is about 0.55% and the recovery is about 94%-107%. The method is accurate, rapid and convenient with satisfactory results.


Subject(s)
Bismuth/analysis , Borohydrides/chemistry , Medicine, Chinese Traditional , Spectrometry, Fluorescence/methods , Bismuth/chemistry , Bismuth/isolation & purification , Drugs, Chinese Herbal/analysis , Drugs, Chinese Herbal/chemistry , Drugs, Chinese Herbal/standards , Hydrochloric Acid/chemistry , Hydrogen Peroxide/chemistry , Hydrogen-Ion Concentration , Nitric Acid/chemistry , Reproducibility of Results , Sulfuric Acids/chemistry