1.
Phys Med Biol ; 68(24)2023 Dec 12.
Article in English | MEDLINE | ID: mdl-37983915

ABSTRACT

Objective.Head motion correction (MC) is an essential process in brain positron emission tomography (PET) imaging. We have used the Polaris Vicra, an optical hardware-based motion tracking (HMT) device, for PET head MC. However, this requires attachment of a marker to the subject's head. Markerless HMT (MLMT) methods are more convenient for clinical translation than HMT with external markers. In this study, we validated the United Imaging Healthcare motion tracking (UMT) MLMT system using phantom and human point source studies, and tested its effectiveness on eight18F-FPEB and four11C-LSN3172176 human studies, with frame-based region of interest (ROI) analysis. We also proposed an evaluation metric, registration quality (RQ), and compared it to a data-driven evaluation method, motion-corrected centroid-of-distribution (MCCOD).Approach.UMT utilized a stereovision camera with infrared structured light to capture the subject's real-time 3D facial surface. Each point cloud, acquired at up to 30 Hz, was registered to the reference cloud using a rigid-body iterative closest point registration algorithm.Main results.In the phantom point source study, UMT exhibited superior reconstruction results than the Vicra with higher spatial resolution (0.35 ± 0.27 mm) and smaller residual displacements (0.12 ± 0.10 mm). In the human point source study, UMT achieved comparable performance as Vicra on spatial resolution with lower noise. Moreover, UMT achieved comparable ROI values as Vicra for all the human studies, with negligible mean standard uptake value differences, while no MC results showed significant negative bias. TheRQevaluation metric demonstrated the effectiveness of UMT and yielded comparable results to MCCOD.Significance.We performed an initial validation of a commercial MLMT system against the Vicra. Generally, UMT achieved comparable motion-tracking results in all studies and the effectiveness of UMT-based MC was demonstrated.


Subject(s)
Image Processing, Computer-Assisted , Positron-Emission Tomography , Humans , Image Processing, Computer-Assisted/methods , Positron-Emission Tomography/methods , Head/diagnostic imaging , Brain/diagnostic imaging , Motion , Phantoms, Imaging , Algorithms , Movement
2.
Magn Reson Med ; 89(2): 605-619, 2023 02.
Article in English | MEDLINE | ID: mdl-36198013

ABSTRACT

PURPOSE: Subject head motion is a major challenge in DWI, leading to image blurring, signal losses, and biases in the estimated diffusion parameters. Here, we investigate a combined application of prospective motion correction and spatial-angular locally low-rank constrained reconstruction to obtain robust, multi-shot, high-resolution diffusion-weighted MRI under substantial motion. METHODS: Single-shot EPI with retrospective motion correction can mitigate motion artifacts and resolve any mismatching of gradient encoding orientations; however, it is limited by low spatial resolution and image distortions. Multi-shot acquisition strategies could achieve higher resolution and image fidelity but increase the vulnerability to motion artifacts and phase variations related to cardiac pulsations from shot to shot. We use prospective motion correction with optical markerless motion tracking to remove artifacts and reduce image blurring due to bulk motion, combined with locally low-rank regularization to correct for remaining artifacts due to shot-to-shot phase variations. RESULTS: The approach was evaluated on healthy adult volunteers at 3 Tesla under different motion patterns. In multi-shot DWI, image blurring due to motion with 20 mm translations and 30° rotations was successfully removed by prospective motion correction, and aliasing artifacts caused by shot-to-shot phase variations were addressed by locally low-rank regularization. The ability of prospective motion correction to preserve the orientational information in DTI without requiring a reorientation of the b-matrix is highlighted. CONCLUSION: The described technique is proved to hold valuable potential for mapping brain diffusivity and connectivity at high resolution for studies in subjects/cohorts where motion is common, including neonates, pediatrics, and patients with neurological disorders.


Subject(s)
Echo-Planar Imaging , Image Interpretation, Computer-Assisted , Adult , Infant, Newborn , Humans , Child , Echo-Planar Imaging/methods , Image Interpretation, Computer-Assisted/methods , Prospective Studies , Retrospective Studies , Diffusion Magnetic Resonance Imaging/methods , Artifacts , Motion , Brain/diagnostic imaging , Image Processing, Computer-Assisted/methods , Algorithms
3.
EJNMMI Phys ; 8(1): 47, 2021 Jun 12.
Article in English | MEDLINE | ID: mdl-34117943

ABSTRACT

BACKGROUND: Integrated whole-body PET/MR technology continues to mature and is now extensively used in clinical settings. However, due to the special design architecture, integrated whole-body PET/MR comes with a few inherent limitations. Firstly, whole-body PET/MR lacks sensitivity and resolution for focused organs. Secondly, broader clinical access of integrated PET/MR has been significantly restricted due to its prohibitively high cost. The MR-compatible PET insert is an independent and removable PET scanner which can be placed within an MRI bore. However, the mobility and configurability of all existing MR-compatible PET insert prototypes remain limited. METHODS: An MR-compatible portable PET insert prototype, dual-panel portable PET (DP-PET), has been developed for simultaneous PET/MR imaging. Using SiPM, digital readout electronics, novel carbon fiber shielding, phase-change cooling, and MRI compatible battery power, DP-PET was designed to achieve high-sensitivity and high-resolution with compatibility with a clinical 3-T MRI scanner. A GPU-based reconstruction method with resolution modeling (RM) has been developed for the DP-PET reconstruction. We evaluated the system performance on PET resolution, sensitivity, image quality, and the PET/MR interference. RESULTS: The initial results reveal that the DP-PET prototype worked as expected in the MRI bore and caused minimal compromise to the MRI image quality. The PET performance was measured to show a spatial resolution ≤ 2.5 mm (parallel to the detector panels), maximum sensitivity = 3.6% at the center of FOV, and energy resolution = 12.43%. MR pulsing introduces less than 2% variation to the PET performance measurement results. CONCLUSIONS: We developed a MR-compatible PET insert prototype and performed several studies to begin to characterize the performance of the proposed DP-PET. The results showed that the proposed DP-PET performed well in the MRI bore and would cause little influence on the MRI images. The Derenzo phantom test showed that the proposed reconstruction method could obtain high-quality images using DP-PET.

4.
Biomed Res Int ; 2019: 8213215, 2019.
Article in English | MEDLINE | ID: mdl-31886254

ABSTRACT

The aim of this study was to validate quantitative performance of a newly released simultaneous positron emission tomography (PET)/magnetic resonance imaging (MRI) scanner, by using MR-based attenuation correction (MRAC), both in phantom study and in patient study. PET/MRI image uniformities of a phantom under different hardware configurations were tested and compared. Thirty patients were examined with 2-deoxy-2-[18F]fluoro-D-glucose (18F-FDG) PET/computed tomography (CT) and subsequent PET/MRI. PET images from PET/MRI were corrected with MRAC (PETMR), CT-based attenuation maps (µ-maps, PETCT), and segmented CT µ-maps (PETCTSeg) derived from PET/CT. Standardized uptake values (SUVs) were compared among the 3 sets of PET in main organs (bone, liver and lung) and in 52 FDG-avid lesions, including soft-tissue lesions and bone lesions. The result showed that PET imaging uniformities of PET/MRI under different configurations were good (<8.8%). The SUV differences among the 3 sets of PET varied with organs and lesion types. In detail, the mean relative differences of SUV between PETMR and PETCT were as follows: -18.8%, bone (SUVmean); -8.0%, liver (SUVmean); -12.2%, lung (SUVmean); -18.1%, bone lesions (SUVmean); -13.3%, bone lesions (SUVmax); -8.2%, soft-tissue lesions (SUVmean); and -7.3%, soft-tissue lesions (SUVmax). The mean relative differences between PETMR and PETCTSeg were as follows: -19.0%, bone (SUVmean); -3.5%, liver (SUVmean); -3.3%, lung (SUVmean); -19.3%, bone lesions (SUVmean); -17.5%, bone lesions (SUVmax); -5.5%, soft-tissue lesions (SUVmean); and -4.4%, soft-tissue lesions (SUVmax). The differences of SUV between PETMR and PETCT were larger than those between PETMR and PETCTSeg, in both soft tissue and soft-tissue lesions (P < 0.001), but not in bone or bone lesions. In conclusion, MRAC in the newly released PET/MR system is accurate in most tissues, with SUV deviations being generally less than 10%, compared to PET/CT. In bone, however, underestimations can be substantial, which may be partially attributed to segmentation of the MR-based µ-maps.


Subject(s)
Algorithms , Magnetic Resonance Imaging , Positron-Emission Tomography , Whole Body Imaging , Adolescent , Aged , Female , Humans , Image Processing, Computer-Assisted , Male , Middle Aged , Multimodal Imaging , Organ Specificity , Phantoms, Imaging , Tomography, X-Ray Computed