Z Orthop Unfall ; 152(1): 33-5, 2014 Feb.
Article in German | MEDLINE | ID: mdl-24578111

ABSTRACT

An 84-year-old female patient was admitted to our emergency department with a periprosthetic fracture of the left femur with an implanted Avantage prosthesis (Biomet®). Conventional X-ray images revealed besides the fracture a lateralisation of the prosthesis stem. An additional CT scan showed a dissociation of the polyethylene inlay from the head of the prosthesis which took place 3 years ago in the course of a closed reposition of a postoperative luxation. This event shows the relevance of a precise analysis of X-ray images after reposition regarding the exact articulation of the different prosthesis components.


Subject(s)
Femoral Fractures/diagnostic imaging , Femoral Fractures/etiology , Hip Dislocation/diagnostic imaging , Hip Dislocation/surgery , Hip Prosthesis/adverse effects , Prosthesis Failure , Aged, 80 and over , Arthroplasty, Replacement, Hip/adverse effects , Device Removal , Diagnosis, Differential , Equipment Failure Analysis , False Negative Reactions , Female , Femoral Fractures/surgery , Hip Dislocation/complications , Humans , Prosthesis Design , Radiography , Treatment Failure