Phys Med Biol ; 58(8): L17-30, 2013 Apr 21.
Article in English | MEDLINE | ID: mdl-23528479

ABSTRACT

A phantom for image quality evaluation of digital mammography systems is presented and compared to the most widely used phantoms in Europe and the US. The phantom contains objects for subjective detection of Landolt rings (four-alternative, forced-choice task) and for objective calculation of signal-difference-to-noise ratios (SDNR), both in a titanium background within a 12-step wedge. Evaluating phantom images corresponding to exposures between 15 and 160 mAs (average glandular dose between 0.2 and 2 mGy), the resulting scores were compared to the scores obtained following the European EPQC and American College of Radiology (ACR) protocols. Scores of the Landolt test equal to 19 and 8.5 and SDNR equal to 20 and 11 were found to be equivalent to the acceptable limiting values suggested by EPQC and ACR. In addition, the Landolt and SDNR tests were shown to take into account the anatomical variations in thickness and tissue density within the breast. The simplified evaluation method presented was shown to be a sensitive, efficient and reliable alternative for image quality evaluation of mammography systems.


Subject(s)
Mammography/instrumentation , Phantoms, Imaging , Titanium , Evaluation Studies as Topic , Polymethyl Methacrylate , Quality Control , Reproducibility of Results