1.
J Mech Behav Biomed Mater ; 49: 310-20, 2015 Sep.
Article in English | MEDLINE | ID: mdl-26057364

ABSTRACT

A synthetic meniscus implant was recently developed for the treatment of patients with mild to moderate osteoarthritis with knee pain associated with medial joint overload. The implant is distinctively different from most orthopedic implants in its pliable construction, and non-anchored design, which enables implantation through a mini-arthrotomy without disruption to the bone, cartilage, and ligaments. Due to these features, it is important to show that the material and design can withstand knee joint conditions. This study evaluated the long-term performance of this device by simulating loading for a total of 5 million gait cycles (Mc), corresponding to approximately five years of service in-vivo. All five implants remained in good condition and did not dislodge from the joint space during the simulation. Mild abrasion was detected by electron microscopy, but µ-CT scans of the implants confirmed that the damage was confined to the superficial surfaces. The average gravimetric wear rate was 14.5 mg/Mc, whereas volumetric changes in reconstructed µ-CT scans point to an average wear rate of 15.76 mm(3)/Mc (18.8 mg/Mc). Particles isolated from the lubricant had average diameter of 15 µm. The wear performance of this polycarbonate-urethane meniscus implant concept under ISO-14243 loading conditions is encouraging.


Subject(s)
Materials Testing , Mechanical Phenomena , Menisci, Tibial , Prostheses and Implants , X-Ray Microtomography , Gravitation , Polycarboxylate Cement , Prosthesis Design , Surface Properties , Urethane
2.
J Mech Behav Biomed Mater ; 29: 42-55, 2014 Jan.
Article in English | MEDLINE | ID: mdl-24055793

ABSTRACT

There are significant potential advantages for restoration of meniscal function using a bio-stable synthetic implant that combines long-term durability with a dependable biomechanical performance resembling that of the natural meniscus. A novel meniscus implant made of a compliant polycarbonate-urethane matrix reinforced with high modulus ultrahigh molecular weight polyethylene fibers was designed as a composite structure that mimics the structural elements of the natural medial meniscus. The overall success of such an implant is linked on its capability to replicate the stress distribution in the knee over the long-term. As this function of the device is directly dependent on its mechanical properties, changes to the material due to exposure to the joint environment and repeated loading could have non-trivial influences on the viscoelastic properties of the implant. Thus, the goal of this study was to measure and characterize the strain-rate response, as well as the viscoelastic properties of the implant as measured by creep, stress relaxation, and hysteresis after simulated use, by subjecting the implant to realistic joint loads up to 2 million cycles in a joint-like setting. The meniscus implant behaved as a non-linear viscoelastic material. The implant underwent minimal plastic deformation after 2 million fatigue loading cycles. Under low compressive loads, the implant was fairly flexible, and able to deform relatively easily (E=120-200 kPa). However as the compressive load applied on the implant was increased, the implant became stiffer (E=3.8-5.2 MPa), to resist deformation. The meniscus implant appears well-matched to the viscoelastic properties of the natural meniscus, and importantly, these properties were found to remain stable and minimally affected by potentially degradative and loading conditions associated with long-term use.


Subject(s)
Materials Testing , Menisci, Tibial , Prostheses and Implants , Viscoelastic Substances , Absorption , Stress, Mechanical , Weight-Bearing