Med Phys ; 35(4): 1494-6, 2008 Apr.
Article in English | MEDLINE | ID: mdl-18491544

ABSTRACT

The authors investigated the peripheral dose reduction for CyberKnife radiosurgery treatments after the installation of a linac shielding upgrade. As in a previous investigation, the authors considered two treatment plans, one for a hypothetical target in the brain and another for a target in the thorax, delivered to an anthropomorphic phantom. The results of the prior investigation showed that the CyberKnife delivered significantly higher peripheral doses than comparable model C Gamma Knife or IMRT treatments. Current measurements, after the linac shielding upgrade, demonstrate that the additional shielding decreased the peripheral dose, expressed as a percentage of the delivered monitor units (MU), by a maximum of 59%. The dose reduction was greatest for cranial-caudal distances from the field edge less than 30 cm, and at these distances, the CyberKnife peripheral dose, expressed as a percentage of the delivered MU, is now comparable to that measured for the other treatment modalities in our previous investigation. For distances between 30 and 70 cm from the field edge, the additional shielding reduced the peripheral dose by between 20% and 55%. At these distances, the CyberKnife peripheral dose remains higher than doses measured in our previous study for the model C Gamma Knife and IMRT.


Subject(s)
Laser Therapy/instrumentation , Particle Accelerators/instrumentation , Radiation Protection/instrumentation , Radiometry/methods , Equipment Design , Equipment Failure Analysis , Laser Therapy/methods , Radiotherapy Dosage
2.
J Appl Clin Med Phys ; 8(2): 26-36, 2007 Apr 11.
Article in English | MEDLINE | ID: mdl-17592464

ABSTRACT

The first aim of the present study was to investigate the feasibility of using fewer beam angles to improve delivery efficiency for the treatment of oropharyngeal cancer (OPC) with inverse-planned intensity-modulated radiation therapy (IP-IMRT). A secondary aim was to evaluate whether the simplified IP-IMRT plans could reduce the indirect radiation dose. The treatment plans for 5 consecutive OPC patients previously treated with a forward-planned IMRT (FP-IMRT) technique were selected as benchmarks for this study. The initial treatment goal for these patients was to deliver 70 Gy to > or = 95% of the planning gross tumor volume (PTV-70) and 59.4 Gy to > or = 95% of the planning clinical tumor volume (PTV-59.4) simultaneously. Each case was re-planned using IP-IMRT with multiple beam-angle arrangements, including four complex IP-IMRT plans using 7 or more beam angles, and one simple IMRT plan using 5 beam angles. The complex IP-IMRT plans and simple IP-IMRT plans were compared to each other and to the FPIMRT plans by analyzing the dose coverage of the target volumes, the plan homogeneity, the dose-volume histograms of critical structures, and the treatment delivery parameters including delivery time and the total number of monitor units (MUs). When comparing the plans, we found no significant difference between the complex IP-IMRT, simple IP-IMRT, and FP-IMRT plans for tumor target coverage (PTV-70: p = 0.56; PTV-59.4: p = 0.20). The plan homogeneity, measured by the mean percentage isodose, did not significantly differ between the IP-IMRT and FP-IMRT plans (p = 0.08), although we observed a trend toward greater inhomogeneity of dose in the simple IP-IMRT plans. All IP-IMRT plans either met or exceeded the quality of the FP-IMRT plans in terms of dose to adjacent critical structures, including the parotids, spinal cord, and brainstem. As compared with the complex IP-IMRT plans, the simple IP-IMRT plans significantly reduced the mean treatment time (maximum probability for four pairwise comparisons: p = 0.0003). In conclusion, our study demonstrates that, as compared with complex IP-IMRT, simple IP-IMRT can significantly improve treatment delivery efficiency while maintaining similar target coverage and sparing of critical structures. However, the improved efficiency does not significantly reduce the total number of MUs nor the indirect radiation dose.


Subject(s)
Models, Biological , Oropharyngeal Neoplasms/radiotherapy , Radiometry/methods , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Conformal/methods , Body Burden , Computer Simulation , Humans , Radiotherapy Dosage , Relative Biological Effectiveness