Anal Chem ; 81(13): 5165-71, 2009 Jul 01.
Article in English | MEDLINE | ID: mdl-19473010

ABSTRACT

We have developed a new protein microarray (ImmunoFlow Protein Platform, IFPP) that utilizes a porous nitrocellulose (NC) membrane with printed spots of capture probes. The sample is pumped actively through the NC membrane, to enhance binding efficiency and introduce stringency. Compared to protein microarrays assayed with the conventional incubation-shaking method the rate of binding is enhanced on the IFPP by at least a factor of 10, so that the total assay time can be reduced drastically without compromising sensitivity. Similarly, the sensitivity can be improved. We demonstrate the detection of 1 pM of C-reactive protein (CRP) in 70 microL of plasma within a total assay time of 7 min. The small sample and reagent volumes, combined with the speed of the assay, make our IFPP also well-suited for a point-of-care/near-patient setting. The potential clinical application of the IFPP is demonstrated by validating CRP detection both in human plasma and serum samples against standard clinical laboratory methods.


Subject(s)
C-Reactive Protein/analysis , Protein Array Analysis/methods , Antibodies/chemistry , Antibodies/immunology , Collodion/chemistry , Humans , Kinetics , Membranes, Artificial , Microscopy, Confocal , Protein Array Analysis/instrumentation