Int J Hyperthermia ; 15(2): 145-56, 1999.
Article in English | MEDLINE | ID: mdl-10323621

ABSTRACT

PURPOSE: The 27 MHz Multi Electrode Current Source (MECS) interstitial hyperthermia system uses probes consisting of multiple independent electrodes, 10-20 mm long, to steer the 3-D power deposition. Seven point thermocouples integrated into the probes provide matching 3-D temperature feedback data. To improve spatial steering the number of independent segments was increased; the feasibility and reliability of four independent electrodes integrated into a single probe were evaluated, with special attention to efficiency and to interference between separate electrodes. METHODS: The contribution of secondary coupling on the apparent electrode impedance and the dependence of cross coupling on the distance between leads, thermocouple and electrodes are computed using simple analytical models. The effect of this secondary coupling was assessed experimentally by comparing power delivery by dual and quadruple electrodes, and by quadruple electrodes in different electrode configurations (segment length 10 or 20 mm) in a nylon catheter in a muscle equivalent medium. RESULTS: Cross coupling with the thermocouple and other electrodes was computed to be of the same magnitude as the primary coupling for a quadruple electrode. Fortunately, this does not affect operation of the electrode, there was no difference in performance between quadruple and dual electrodes, and the output power was effectively independent of the electrode configuration. CONCLUSION: Quadruple MECS electrodes for improved 3-D power control are feasible.


Subject(s)
Electrodes , Hyperthermia, Induced/instrumentation
2.
Phys Med Biol ; 43(1): 139-53, 1998 Jan.
Article in English | MEDLINE | ID: mdl-9483628

ABSTRACT

The multi-electrode current source (MECS) interstitial hyperthermia system which is used for treatment of cancer, employs segmented electrodes inserted in plastic tubes implanted in the treatment volume. The mean power deposition of the individual electrodes is controlled by varying the duty cycle of the RF signal applied to the electrodes, using thermocouples inside the electrodes for thermometry. A non-zero loss angle results in self-heating of the catheter. The thermal influence of self-heating was investigated and an analysis of the measurement of temperatures inside the catheter during and after heating is presented. Analytical models and a high-resolution numerical model were used for the calculation of steady state and transient distributions, respectively. The model results are compared with experimental data obtained in a muscle equivalent phantom. Results indicate that there is no difference between temperature inside and outside the catheter when using lossless catheter materials (e.g. PE and PTFE). Self-heating in the catheter wall has an adverse effect on the uniformity of the stationary temperature distribution and the reliability of temperature measurement with internal thermometry. These problems remain within acceptable limits for mildly lossy materials; the difference between the temperature inside and outside is only 6% when using low-loss Nylon. Analysis of the thermal decay after power-off shows that low-loss materials allow more time to obtain an accurate estimate of the tissue temperature at the catheter wall during power-on. This effect is enhanced by the presence of minute air layers in the applicator. Distortion of temperature gradients along the catheter was also investigated. Key factors are the thermal conduction across the catheter wall, and especially the presence of minute layers of air between consecutive layers of the probe. The distortion extends less than two millimetres, which is acceptable. The simulation results are compatible with measurements in phantoms and show that, if the proper choice of materials is made, the MECS applicator answers our expectations and that the temperature measurement inside the catheter can be used for direct feedback treatment control.


Subject(s)
Electrodes , Hyperthermia, Induced/instrumentation , Biophysical Phenomena , Biophysics , Catheterization , Electric Conductivity , Humans , Hyperthermia, Induced/methods , Models, Biological , Temperature
3.
Phys Med Biol ; 42(7): 1451-60, 1997 Jul.
Article in English | MEDLINE | ID: mdl-9253052

ABSTRACT

We have developed a thermal model in which blood vessels are described as geometrical objects, 3D curves with associated diameters. Here the behaviour of the model is examined for low resolutions compared with the vessel diameter and for strongly curved vessels. The tests include a single straight vessel and vessels describing the path of a helix embedded in square tissue blocks. The tests show the excellent behaviour of our discrete vessel implementation.


Subject(s)
Blood Vessels/physiology , Models, Biological , Models, Structural , Hot Temperature , Humans , Mathematics , Models, Cardiovascular , Reproducibility of Results
4.
Int J Hyperthermia ; 13(4): 365-85, 1997.
Article in English | MEDLINE | ID: mdl-9278767

ABSTRACT

The quality of temperature distributions that can be generated with the Multi Electrode Current Source (MECS) interstitial hyperthermia (IHT) system, which allows 3D control of the temperature distribution, has been investigated. For the investigations, computer models of idealised anatomies containing discrete vessels, were used. A 7-catheter hexagonal implant geometry with a nearest neighbour distance of 15 mm was used. In each interstitial catheter with a diameter of 2.1 mm a number of 1 up to 4 electrodes were placed along an 'active section' with a length of 50 mm. The electrode segments had lengths of 50, 20, 12 and 9 mm respectively. Both single vessel and vessel network situations were analysed. This study shows that even in situations with discrete vasculature and perfusion heterogeneity it remains possible to obtain satisfactory temperature distributions with the MECS IHT system. Due to its 3D spatial control the temperature homogeneity in the implant can be made quite satisfactory.


Subject(s)
Hyperthermia, Induced/methods , Blood Vessels/anatomy & histology , Blood Vessels/physiology , Computer Simulation , Electrodes , Evaluation Studies as Topic , Humans , Hyperthermia, Induced/instrumentation , Hyperthermia, Induced/statistics & numerical data , Models, Anatomic , Models, Cardiovascular , Temperature
5.
Phys Med Biol ; 42(4): 637-50, 1997 Apr.
Article in English | MEDLINE | ID: mdl-9127442

ABSTRACT

The 27 MHz Multi Electrode Current Source (MECS) interstitial hyperthermia system uses segmented electrodes, 10-20 mm long, to steer the 3D power deposition. This power control at a scale of 1-2 cm requires detailed and accurate temperature feedback data. To this end seven-point thermocouples are integrated into the probes. The aim of this work was to evaluate the feasibility and reliability of integrated thermometry in the 27 MHz MECS system, with special attention to the interference between electrode and thermometry and its effect on system performance. We investigated the impact of a seven-sensor thermocouple probe (outer diameter 150 microns) on the apparent impedance and power output of a 20 mm dual electrode (O.D. 1.5 mm) in a polyethylene catheter in a muscle equivalent medium (sigma 1 = 0.6 S m-1). The cross coupling between electrode and thermocouple was found to be small (1-2 pF) and to cause no problems in the dual-electrode mode, and only minimal problems in the single-electrode mode. Power loss into the thermometry system can be prevented using simple filters. The temperature readings are reliable and representative of the actual tissue temperature around the electrode. Self-heating effects, occurring in some catheter materials, are eliminated by sampling the temperature after a short power-off interval. We conclude that integrated thermocouple thermometry is compatible with 27 MHz capacitively coupled interstitial hyperthermia. The performance of the system is not affected and the temperatures measured are a reliable indication of the maximum tissue temperatures.


Subject(s)
Hyperthermia, Induced/instrumentation , Electrodes , Equipment Design , Hyperthermia, Induced/methods , Models, Theoretical , Temperature
6.
Int J Hyperthermia ; 13(6): 607-19, 1997.
Article in English | MEDLINE | ID: mdl-9421742

ABSTRACT

Multi electrode current source interstitial hyperthermia (MECS-IHT) employs individually controlled, 27 MHz radiofrequency electrodes inserted into plastic brachytherapy catheters. In order to get a firm understanding of the physical behaviour of the electrodes and to verify the current source approximation in our hyperthermia treatment planning system we have investigated (1) the electrical properties of the electrode-catheter-tissue system, and (2) the impact of inhomogeneity of the electrical properties of the tissue in the vicinity of the electrodes. The results validate the use of the ideal current source approximation in the treatment planning SAR model. The models predict the presence of a significant heat source inside the electrode wall when lossy catheter materials are used, producing a conductive heating component in addition to the SAR in the tissue. For a given catheter spacing this conductive component will produce a more heterogeneous temperature distribution. Thus, low-loss catheter materials like polyethylene and Teflon are recommended. The SAR is highly localized near the catheter. Calculations concerning a fat-muscle interface show that the SAR is higher in the fatty tissue than in the muscle tissue; 3D SAR control by individually controlled electrode segments is essential in such a situation.


Subject(s)
Hyperthermia, Induced/instrumentation , Models, Theoretical , Animals , Electrodes , Humans , Hyperthermia, Induced/methods
7.
IEEE Trans Biomed Eng ; 43(10): 1038-45, 1996 Oct.
Article in English | MEDLINE | ID: mdl-9214821

ABSTRACT

A three-dimensional (3-D) model is presented for the calculation of the specific absorption rate (SAR) in human tissue during current source interstitial hyperthermia. The model is capable of millimeter resolution and can cope with irregular implants in heterogeneous tissue. The SAR distribution is calculated from the electrical potential. The potential distribution is determined by the dielectric properties of the tissue and by the electrode configuration. The dielectric properties and the current injection of the electrodes are represented on a 3-D uniform grid. The calculated potential at an electrode current injection point is not the actual electrode potential at that point. To estimate this potential a grid independent representation of an electrode together with an analytical solution in the neighborhood of the electrode are used. The calculated potential on the electrode surface is used to estimate the electrode impedance. The tissue implementation is validated by comparing calculated distributions with analytical solutions. The electrode implementation is verified by comparing different discretizations of an electrode configuration and by comparing numerically calculated electrode impedances with analytically calculated impedances.


Subject(s)
Hyperthermia, Induced , Models, Biological , Absorption , Brain/anatomy & histology , Electric Conductivity , Electric Impedance , Electrodes , Electromagnetic Fields , Equipment Design , Humans , Hyperthermia, Induced/instrumentation , Magnetic Resonance Imaging , Surface Properties
8.
Phys Med Biol ; 41(3): 429-44, 1996 Mar.
Article in English | MEDLINE | ID: mdl-8778824

ABSTRACT

The quality of temperature distributions that can be generated with the multi-electrode current source (MECS) interstitial hyperthermia system, which allows 3D control of the spatial SAR distribution, has been investigated. For the investigations, computer models of idealized anatomies were used. These anatomical models did not contain discrete vessels. Binary-media anatomies, containing media interfaces oriented parallel, perpendicular or oblique with respect to the long axis of the implant, represent simple anatomies which can be encountered in the clinic. The implant volume was about 40 cm3. A seven-catheter hexagonal implant geometry with a nearest-neighbor distance of 15 mm was used. In each interstitial probe between one and four electrodes with a diameter of 2.1 mm were placed along an "active section' with a length of 50 mm. The electrode segments had lengths of 50, 20, 12 and 9 mm. This study shows that even with high contrasts in electrical and thermal conductivity in the implant it remains possible to obtain satisfactory temperature distributions with the MECS system. Due to its 3D spatial control the temperature homogeneity in the implant can be made quite satisfactory, with T10-T90 of the order of 2-3 K. Treatment planning must ensure that the placement of the current source electrodes is compatible with the media configuration.


Subject(s)
Hyperthermia, Induced/methods , Biophysical Phenomena , Biophysics , Computer Simulation , Electrodes , Humans , Hyperthermia, Induced/standards , Hyperthermia, Induced/statistics & numerical data , Models, Anatomic , Neoplasms/therapy , Quality Control , Temperature