1.
Rev. bras. anal. clin ; 52(4): 376-382, 20201230. tab
Article in Portuguese | LILACS | ID: biblio-1248338

ABSTRACT

Objetivo: Avaliar o desempenho do equipamento Mindray® BC-5380 através da métrica sigma. Métodos: Os parâmetros incluídos neste estudo são leucócitos, eritrócitos, hemoglobina, hematócrito, volume corpuscular médio, hemoglobina corpuscular média, concentração de hemoglobina corpuscular média, amplitude de distribuição dos eritrócitos e plaquetas. Os cálculos foram realizados a partir de dados provenientes da rotina de controle de qualidade do laboratório, determinando-se a inexatidão, imprecisão, o erro total analítico e a métrica sigma dos parâmetros. Resultados: Valores aceitáveis de Sigma (>3) calculados com base nas especificações da qualidade CLIA foram encontrados para todos os parâmetros, exceto hematócrito. Calculando-se a métrica com base em especificações de variação biológica, hematócrito, volume corpuscular médio e hemoglobina corpuscular média apresentam desempenho inaceitável em pelo menos algum nível de controle, e a concentração de hemoglobina corpuscular média apresenta resultado inaceitável em todos os níveis. A contagem de leucócitos totais foi o único parâmetro com desempenho excelente em todos os níveis de controle, perante todas as especificações de qualidade aplicadas. Conclusão: A utilização da métrica sigma na avaliação do desempenho analítico permite identificar falhas não detectadas pelo controle de qualidade convencional, sendo um importante recurso para a melhoria contínua da qualidade.


Objective: To evaluate the performance of the Mindray® BC-5380 equipment using the sigma metric. Methods: The parameters included in this study are leukocytes, erythrocytes, hemoglobin, hematocrit, average corpuscular volume, mean corpuscular hemoglobin, mean corpuscular hemoglobin concentration, red cell distribution width and platelets. The calculations were performed based on data from the laboratory's quality control routine, determining inaccuracy, imprecision, total analytical error and the sigma metric of the parameters. Results: Acceptable values of Sigma (> 3) calculated based on CLIA specifications were found for all parameters, except hematocrit Calculating the metric based on specifications of biological variation, hematocrit, mean corpuscular volume and mean corpuscular hemoglobin present unacceptable performance in at least one level of control, and the mean corpuscular hemoglobin concentration shows unacceptable results at all levels. The total leukocyte count was the only parameter with excellent performance at all levels of control, with the quality specifications applied. Conclusion: The use of the sigma metric in the evaluation of analytical performance allows to identify flaws not detected by conventional quality control, being an important resource for continuous quality improvement.


Subject(s)
Quality Control , Total Quality Management , Hematology
2.
J. bras. patol. med. lab ; 52(1): 11-16, Jan.-Feb. 2016. tab
Article in English | LILACS | ID: lil-775602

ABSTRACT

ABSTRACT Introduction: Clinical laboratory results influence more than 60% of medical decisions, impacting on disease prevention, diagnosis, and treatment. The use of quality indicators (QIs) is fundamental for quality assurance, once it allows monitoring the process and directs the taking of corrective actions. Objective: Classify the biological sample recollection requests for QIs identification in a clinical laboratory of Distrito Federal, Brazil. Material and methods: Data about the requests made in the biennium 2013-2014 were gathered and analyzed. Results and discussion: Among the 304,361 samples received, 1,914 (0.62%) had a request for recollection made in accordance with laboratory criteria. Most orders originated in the pre-analytical phase (57.7%). The most frequent reason for sample rejection was result confirmation (40.7%), followed by insufficient sample (21.9%), coagulated sample (18.1%) and hemolyzed sample (11, 9%). The hematology sector was responsible for most recollection requests (43.6%), followed by the biochemistry (29%) and the immunology (25.7%) ones. The laboratory emergency department accounted for only 0.1%. Orders were mostly placed by the outpatient clinic (40.7%), emergency (30.4%), and internal medicine (12.4%) departments. The percentage of orders is low, but does not exclude the need to reach lower rates. Underreporting was detected in the emergency sector, which indicates need for improvement in information registration. Conclusion: The numbers mentioned were selected as IQs for the pre-analytical phase, serving as guidelines for future actions taken by the team.


RESUMO Introdução: Os resultados dos testes laboratoriais influenciam mais de 60% das decisões médicas, impactando na prevenção, no diagnóstico e no tratamento de doenças. O uso de indicadores de qualidade (IQs) é fundamental para a garantia da qualidade, pois permite monitoramento do processo e direciona a implementação de ações corretivas. Objetivo: Classificar pedidos de recoleta de amostras biológicas para identificação de IQs em um laboratório de análises clínicas do Distrito Federal. Material e métodos: Foram coletados e analisados dados acerca de pedidos de recoleta feitos no biênio 2013-2014. Resultados e discussão: Das 304.361 amostras registradas, 1.914 (0,62%) tiveram solicitação de recoleta de acordo com os critérios do laboratório. A maioria dos pedidos teve origem na fase pré-analítica (57,7%). O motivo de rejeição mais frequente foi confirmação de resultado (40,7%), seguido por amostra insuficiente (21,9%), amostra coagulada (18,1%) e amostra hemolisada (11,9%). O setor de hematologiafoi o responsável pelo maior número de pedidos de recoletas (43,6%), seguido pelo de bioquímica (29%) e o de imunologia (25,7%). O setor de emergência do laboratório representou apenas 0,1%. Quanto à procedência, a maioria das recoletas foi solicitada pelo ambulatório (40,7%), pelo pronto-socorro (30,4%) e pela clínica médica (12,4%). A porcentagem de pedidos de recoleta é baixa, porém não exclui a necessidade de busca por menores índices. Sugere-se que tenha ocorrido subnotificação no setor de emergência do laboratório, o que aponta necessidade de melhoria no registro de informações. Conclusão: Os números destacados foram selecionados como IQs para a fase pré-analítica, servindo como norteadores para as futuras ações corretivas efetuadas pela equipe.