Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 11 de 11
Filter
1.
Phys Med Biol ; 69(6)2024 Mar 13.
Article in English | MEDLINE | ID: mdl-38408387

ABSTRACT

Objective. Real-time MRgRT uses 2D-cine imaging for target tracking and motion evaluation. Rotation of gantry inducedB0off-resonance, resulting in image artifacts and imaging isocenter-shift precluding MR-guided arc therapy. Standard MRI phantoms designed for higher resolution images face challenges when low-resolution cine imaging is needed to achieve high frame rates. This work aimed to examine the spatial accuracy including geometric distortion and isocenter shift in real-time during gantry rotation on a 0.35 T MR-Linac using the concentric Cine imaging quality assurance (QA) phantom and its associated image analysis software.Approach. The Cine imaging QA phantom consists of two concentric shells of low-T1mineral oil and a central alignment structure. The phantom was scanned on three different MRI systems; 0.55 T Siemens Free.Max, 1.5 T Philips Ingenia, and 0.35 T ViewRay MRIdian MR-Linac using 2D balanced steady-state free precession (bSSFP) imaging sequence. In addition, bSSFP cine MRI with the banding artifact correction was tested on 0.35 T ViewRay MR-Linac. Images from the MR-Linac were acquired with the Linac gantry stationary and rotating from gantry 300°â†’ 0° and vice versa. Three orthogonal image planes were scanned excluding the 1.5 T Philips Ingenia, where only the axial plane was scanned. The image analysis software calculated the distortion values as well as the isocenter position for each cine frame.Main results. The geometric distortion of cine imaging on MRIs and MR-Linac at gantry stationary are within 1 mm while the substantial geometric distortion of 2 and 2.2 mm were observed on 0.35 T MR-Linac while rotating the gantry clockwise (300°â†’ 0°) and counterclockwise 0°â†’ 300° respectively. The average imaging isocenter shift was 0.1 mm for both MRIs and the static gantry and imaging isocenter shift of ≤1.5 mm was observed during the gantry rotation. The imaging isocenter shift decreased by 1 ± 0.2 mm clockwise and counterclockwise withB0compensation.Significance. The concentric Cine imaging QA phantom and its associated software effectively demonstrate the image distortion on real-time cine imaging on regular MRIs and 0.35 T MR-Linac. The results of significant geometric distortion with a rotating gantry in the MR-Linac system require further investigation to alleviate the extent of the image distortion.


Subject(s)
Image Processing, Computer-Assisted , Particle Accelerators , Image Processing, Computer-Assisted/methods , Software , Phantoms, Imaging , Magnetic Resonance Imaging/methods
2.
Med Phys ; 49(4): 2355-2365, 2022 Apr.
Article in English | MEDLINE | ID: mdl-35100445

ABSTRACT

PURPOSE: To describe the creation process of a new breast phantom specifically designed to monitor quality control (QC) metrics consistency over several months in digital breast tomosynthesis (DBT). METHODS: The semi-anthropomorphic Tomomam® phantom was designed and evaluated twice monthly on a single Hologic Selenia Dimensions® unit over 5 months. The phantom is manufactured in a one-piece epoxy resin homogeneous material as the basis for manufacturing, simulating breast tissue as 50% equivalent glandular (GL)/50% equivalent adipose (AD) and compressed thickness of 60 mm. The distribution of test objects on different planes inside the phantom should allow the quantification of 10 image quality metrics: reproducibility, signal difference-to-noise ratio (SDNR), geometric distortions in the plane, missing or added tissue at chest wall, at the top and bottom of images stack and lateral sides, in-plane homogeneity, image scoring, artifact spread function (ASF), geometric distortions in the volume. SDNR was quantified according to GL and AD tissues. Tolerance criteria per parameter were described to analyze results over the study time. RESULTS: Mean scores were equal to 15.4, 15.0, and 11.6 for masses, microcalcifications, and fibers, respectively. A large difference between GL and AD tissues for SDNR metrics was noted over the study time: the best results were obtained from GL tissues. Both geometric distortions and local homogeneity in the plane conformed to expected values. The mean volume value of the triangular prism was 11.3% greater than the expected value due to a reconstruction height equal to 66 mm instead of 60 mm. CONCLUSIONS: In this study, we monitored several QC metrics discriminating GL and AD tissues by using a new breast phantom developed by us. The preliminary clinical tests demonstrated that the Tomomam® phantom could be used to reliably and efficiently track 10 QC metrics with a single acquisition. More data need to be acquired to refine tolerance criteria for some metrics.


Subject(s)
Breast , Mammography , Breast/diagnostic imaging , Mammography/methods , Phantoms, Imaging , Radiographic Image Enhancement/methods , Reproducibility of Results , Signal-To-Noise Ratio
3.
Biomed Phys Eng Express ; 8(1)2021 12 09.
Article in English | MEDLINE | ID: mdl-34844222

ABSTRACT

The RUBY head phantom in combination with the System QA insert MultiMet can be used for simultaneous point dose measurements at an isocentric and two off-axis positions. This study investigates the suitability of the system for systematic integral end-to-end testing of single-isocenter multiple target stereotactic treatments. Several volumetric modulated arc therapy plans were optimized on a planning CT of the phantom positioned in a stereotactic mask on the stereotactic treatment board. The plans were created for three artificial spherical target volumes centred around the measurement positions in the MultiMet insert. Target diameters between 5 and 40 mm were investigated. Coplanar and non-coplanar plans were optimized using the collapsed cone algorithm of the Oncentra Masterplan treatment planning system and recalculated with the Monte Carlo algorithm of the Monaco treatment planning system. Measurements were performed at an Elekta Synergy linear accelerator. The head phantom was positioned according to clinical workflow comprising immobilization and CBCT imaging. Simultaneous point dose measurements at all target positions were performed with three PinPoint 3D chambers (type 31022) as well as three microDiamond detectors (type 60019) and compared to the treatment planning system calculations. Furthermore, the angular dependence of the detector response was investigated to estimate the associated impact on the measured point dose values. Considering all investigated plans, PTV diameters and positions, the point doses calculated with the Monaco treatment planning system and the microDiamond measurements differed within 3.5%, whereas the PinPoint 3D showed differences of up to 6.9%. Point dose differences determined in comparison to the Oncentra Masterplan dose calculations were larger. The RUBY system was shown to be suitable for end-to-end testing of complex treatment scenarios such as single-isocenter multiple target plans.


Subject(s)
Radiosurgery , Radiotherapy Planning, Computer-Assisted , Particle Accelerators , Phantoms, Imaging , Radiosurgery/methods , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted/methods
4.
J Appl Clin Med Phys ; 21(12): 84-89, 2020 Dec.
Article in English | MEDLINE | ID: mdl-33136313

ABSTRACT

PURPOSE: This study aimed to develop a high-dose-rate brachytherapy (HDR-BT) quality assurance (QA) tool for verification of source positions, and to report on its effectiveness. METHODS: We fabricated a cuboid phantom measuring 30 × 30×3 cm3 with spaces to embed Fletcher-Williamson tandem and ovoid applicators. Lead-based, cylindrically shaped radiopaque markers, which scatter radiation and blacken the Gafchromic® RTQA2 films placed on the applicators, were inserted into the phantom to determine the applicator tip and reference source positions. A three-dimensional image-guided brachytherapy (3D-IGBT) plan was generated, and the source positions on the film and radiation treatment planning system (RTPS) were verified with the tool. Source position errors were evaluated as the distance in the applicator axis direction between the source position and the center position of two radiopaque marker pairs. RESULTS: Source position errors on the film and RTPS were in good agreement with one another and were all within 0.5 mm for all applicators. Offset values of each applicator were in good agreement with the value determined in treatment planning (6 mm). The expanded measurement uncertainty of our QA tool was estimated to be 0.87 mm, with a coverage factor k of 2. CONCLUSIONS: Our new HDR-BT QA tool developed for comprehensive source position verification will be useful for cross checking actual source positions and planned source positions on the RTPS.


Subject(s)
Brachytherapy , Humans , Imaging, Three-Dimensional , Phantoms, Imaging , Radiotherapy Dosage , Uncertainty
5.
Radiol Phys Technol ; 13(2): 195-200, 2020 Jun.
Article in English | MEDLINE | ID: mdl-32078138

ABSTRACT

Short and semi-automated quality assurance (QA) programs are becoming one of the most popular and highly demanding tasks in radiotherapy. The current research investigates the accuracy of a four degrees of freedom (4DoF) medical linear accelerator couch positioning with a fast and accurate method based on images acquired using an electronic portal imaging device (EPID). An accurate EPID QA phantom and a proper in-house code were used. A Siemens medical linear accelerator equipped with an a-Si EPID was used to acquire portal images. For verifying the mechanical performance of the EPID positioning, EPID sensitivity, and accuracy of the code response from the image processing point of view were investigated. To characterize the results, three deviations in the phantom positioning were deliberately created. The translational and rotational displacements of the treatment couch were then evaluated. The loading effect on the treatment couch was then investigated. The results of prerequisite tests, including the mechanical performance of the EPID, and the sensitivity and accuracy of the recognition codes, were assessed. The results were found to be within the tolerance range reported at AAPM TG-142. The mean deviations of the tests between expected and measured displacements by 4DoF treatment couch were found to be 0.13° ± 0.11°, 0.12 ± 0.17 mm, 0.17 ± 0.13 mm, and 0.04 ± 0.09 mm for rotational, longitudinal, lateral, and vertical shifts, respectively. The results showed that the proposed method is a reliable and fast approach to find the uncertainties occurring intreatment couch positioning.


Subject(s)
Electrical Equipment and Supplies , Particle Accelerators
6.
Med Phys ; 47(2): 604-613, 2020 Feb.
Article in English | MEDLINE | ID: mdl-31808949

ABSTRACT

PURPOSE: The purpose of this paper was to design, manufacture, and evaluate a tissue equivalent, dual magnetic resonance/computed tomography (MR/CT) visible anthropomorphic head and neck (H&N) phantom. This phantom was specially designed as an end-to-end quality assurance (QA) tool for MR imaging guided radiotherapy (MRIgRT) systems participating in NCI-sponsored clinical trials. METHOD: The MRIgRT H&N phantom was constructed using a water-fillable acrylic shell and a custom insert that mimics an organ at risk (OAR) and target structures. The insert consists of a primary and secondary planning target volume (PTV) manufactured of a synthetic Clear Ballistic gel, an acrylic OAR and surrounding tissue fabricated using melted Superflab. Radiochromic EBT3 film and thermoluminescent detectors (TLDs) were used to measure the dose distribution and absolute dose, respectively. The phantom was evaluated by conducting an end-to-end test that included: imaging on a GE Lightspeed CT simulator, planning on Monaco treatment planning software (TPS), verifying treatment setup with MR, and irradiating on Elekta's 1.5 T Unity MR linac system. The phantom was irradiated three times using the same plan to determine reproducibility. Three institutions, equipped with either ViewRay MRIdian 60 Co or ViewRay MRIdian Linac, were used to conduct a feasibility study by performing independent end-to-end studies. Thermoluminescent detectors were evaluated in both reproducibility and feasibility studies by comparing ratios of measured TLD to reported TPS calculated values. Radiochromic film was used to compare measured planar dose distributions to expected TPS distributions. Film was evaluated by using an in-house gamma analysis software to measure the discrepancies between film and TPS. RESULTS: The MRIgRT H&N phantom on the Unity system resulted in reproducible TLD doses (SD < 1.5%). The measured TLD to calculated dose ratios for the Unity system ranged from 0.94 to 0.98. The Viewray dose result comparisons had a larger range (0.95-1.03) but these depended on the TPS dose calculations from each site. Using a 7%/4 mm gamma analysis, Viewray institutions had average axial and sagittal passing rates of 97.3% and 96.2% and the Unity system had average passing rates of 97.8% and 89.7%, respectively. All of the results were within the Imaging and Radiation Oncology Core in Houston (IROC-Houston) standard credentialing criteria of 7% on TLDs, and >85% of pixels passing gamma analysis using 7%/4 mm on films. CONCLUSIONS: An MRIgRT H&N phantom that is tissue equivalent and visible on both CT and MR was developed. The results from initial reproducibility and feasibility testing of the MRIgRT H&N phantom using the tested MGIgRT systems suggests the phantom's potential utility as a credentialing tool for NCI-clinical trials.


Subject(s)
Head/diagnostic imaging , Magnetic Resonance Imaging/instrumentation , Neck/diagnostic imaging , Tomography, X-Ray Computed/instrumentation , Clinical Trials as Topic , Equipment Design , Feasibility Studies , Head/anatomy & histology , Head/radiation effects , Humans , Neck/anatomy & histology , Neck/radiation effects , Phantoms, Imaging , Quality Control , Radiotherapy, Image-Guided
7.
Med Phys ; 46(11): 5124-5133, 2019 Nov.
Article in English | MEDLINE | ID: mdl-31506963

ABSTRACT

PURPOSE: To design, manufacture, and evaluate a dynamic magnetic resonance imaging/computed tomography (MRI/CT)-compatible anthropomorphic thorax phantom used to credential MR image-guided radiotherapy (MRIgRT) systems participating in NCI-sponsored clinical trials. METHOD: The dynamic anthropomorphic thorax phantom was constructed from a water-fillable acrylic shell that contained several internal structures representing radiation-sensitive organs within the thoracic region. A custom MR/CT visible cylindrical insert was designed to simulate the left lung with a centrally located tumor target. The surrounding lung tissue was constructed from a heterogeneous in-house mixture using petroleum jelly and miniature (2-4 mm diameter) styrofoam balls and the tumor structure was manufactured from liquid PVC plastic. An MR conditional pneumatic system was developed to allow the MRIgRT insert to move in similar inhale/exhale motions. TLDs and radiochromic EBT3 film were inserted into the phantom to measure absolute point doses and dose distributions, respectively. The dynamic MRIgRT thorax phantom was evaluated through a reproducibility study and a feasibility study. Comprehensive end-to-end examinations were done where the phantom was imaged on a CT, an IMRT treatment plan was created and an MR image was captured to verify treatment setup. Then, the phantom was treated on an MRIgRT system. The reproducibility study evaluated how well the phantom could be reproduced in an MRIgRT system by irradiating three times on an Elekta's 1.5 T Unity system. The phantom was shipped to three independent institutions and was irradiated on either an MRIdian cobalt-60 (60 Co) or an MRIdian linear accelerator system. Treatment evaluations used TLDs and radiochromic film to compare the planned treatment reported on the treatment planning software against the measured dose on the dosimeters. RESULTS: The phantom on the Unity system had reproducible TLD doses measurements (SD < 1.5%). The measured TLD to calculated dose ratios from the reproducibility and feasibility studies ranged from 0.93 to 1.01 and 0.96 to 1.03, respectively. Using a 7%/5 mm gamma analysis criteria, the reproducibility and feasibility studies resulted in an average passing rate of 93.3% and 96.8%, respectively. No difference was noted in the results between the MRIdian 60 Co and MRIdian 6 MV linac delivery to the phantom and all treatment evaluations were within IROC-Houston's acceptable criterion. CONCLUSIONS: A dosimetrically tissue equivalent, CT/MR visible, motion-enabled anthropomorphic MRIgRT thorax phantom was constructed to simulate a lung cancer patient and was evaluated as an appropriate NIH credentialing tool used for MRIgRT systems.


Subject(s)
Magnetic Resonance Imaging , Phantoms, Imaging , Radiotherapy, Image-Guided/instrumentation , Thorax/diagnostic imaging , Feasibility Studies , Humans , Lung/diagnostic imaging , Quality Control , Radiometry , Reproducibility of Results , Tomography, X-Ray Computed
8.
Med Dosim ; 42(3): 238-246, 2017.
Article in English | MEDLINE | ID: mdl-28645509

ABSTRACT

In a radiotherapy center, daily quality assurance (QA) measurements are performed to ensure that the equipment can be safely used for patient treatment on that day. In a pencil beam scanning (PBS) proton therapy center, spot positioning, spot size, range, and dose output are usually verified every day before treatments. We designed, built, and tested a new, reliable, sensitive, and inexpensive phantom, coupled with an array of ionization chambers, for daily QA that reduces the execution times while preserving the reliability of the test. The phantom is provided with 2 pairs of wedges to sample the Bragg peak at different depths to have a transposition on the transverse plane of the depth dose. Three "boxes" are used to check spot positioning and delivered dose. The box thickness helps spread the single spot and to fit a Gaussian profile on a low resolution detector. We tested whether our new QA solution could detect errors larger than our action levels: 1 mm in spot positioning, 2 mm in range, and 10% in spot size. Execution time was also investigated. Our method is able to correctly detect 98% of spots that are actually in tolerance for spot positioning and 99% of spots out of 1 mm tolerance. All range variations greater than the threshold (2 mm) were correctly detected. The analysis performed over 1 month showed a very good repeatability of spot characteristics. The time taken to perform the daily quality assurance is 20 minutes, a half of the execution time of the former multidevice procedure. This "in-house build" phantom substitutes 2 very expensive detectors (a multilayer ionization chamber [MLIC] and a strip chamber, reducing by 5 times the cost of the equipment. We designed, built, and validated a phantom that allows for accurate, sensitive, fast, and inexpensive daily QA procedures in proton therapy with PBS.


Subject(s)
Phantoms, Imaging , Proton Therapy , Humans , Quality Assurance, Health Care
9.
Phys Med ; 31(5): 529-35, 2015 Jul.
Article in English | MEDLINE | ID: mdl-25964129

ABSTRACT

PURPOSE: To develop and test the suitability and performance of a comprehensive quality assurance (QA) phantom for the Small Animal Radiation Research Platform (SARRP). METHODS AND MATERIALS: A QA phantom was developed for carrying out daily, monthly and annual QA tasks including: imaging, dosimetry and treatment planning system (TPS) performance evaluation of the SARRP. The QA phantom consists of 15 (60 × 60 × 5 mm(3)) kV-energy tissue equivalent solid water slabs. The phantom can incorporate optically stimulated luminescence dosimeters (OSLD), Mosfet or film. One slab, with inserts and another slab with hole patterns are particularly designed for image QA. RESULTS: Output constancy measurement results showed daily variations within 3%. Using the Mosfet in phantom as target, results showed that the difference between TPS calculations and measurements was within 5%. Annual QA results for the Percentage depth dose (PDD) curves, lateral beam profiles, beam flatness and beam profile symmetry were found consistent with results obtained at commissioning. PDD curves obtained using film and OSLDs showed good agreement. Image QA was performed monthly, with image-quality parameters assessed in terms of CBCT image geometric accuracy, CT number accuracy, image spatial resolution, noise and image uniformity. CONCLUSIONS: The results show that the developed QA phantom can be employed as a tool for comprehensive performance evaluation of the SARRP. The study provides a useful reference for development of a comprehensive quality assurance program for the SARRP and other similar small animal irradiators, with proposed tolerances and frequency of required tests.


Subject(s)
Cone-Beam Computed Tomography/instrumentation , Phantoms, Imaging , Radiometry/instrumentation , Radiotherapy Planning, Computer-Assisted/instrumentation , Animals , Quality Control
10.
Article in Korean | WPRIM (Western Pacific) | ID: wpr-115853

ABSTRACT

Radiation treatment for skin cancer has recently increased in tomotherapy. It was reported that required dose could be delivered with homogeneous dose distribution to the target without field matching using electron and photon beam. Therapeutic beam of tomotherapy, however, has several different physical characteristic and irradiation of helical beam is involved in the mechanically dynamic factors. Thus verification of skin dose is requisite using independent tools with additional verification method. Modified phantom for dose measurement was developed and skin dose verification was performed using inserted thermoluminescent dosimeters (TLDs) and GafChromic EBT films. As the homogeneous dose was delivered to the region including surface and 6 mm depth, measured dose using films showed about average 2% lower dose than calculated one in treatment planning system. Region indicating about 14% higher and lower absorbed dose was verified on measured dose distribution. Uniformity of dose distribution on films decreased as compared with that of calculated results. Dose variation affected by inhomogeneous material, Teflon, little showed. In regard to the measured dose and its distribution in tomotherapy, verification of skin dose through measurement is required before the radiation treatment for the target located at the curved surface or superficial depth.


Subject(s)
Electrons , Polytetrafluoroethylene , Radiotherapy, Intensity-Modulated , Skin , Skin Neoplasms
11.
Article in Korean | WPRIM (Western Pacific) | ID: wpr-7198

ABSTRACT

In this study, we developed the protopype of QA phantom for image QA including an additional component for image based radiation treatment system. The new phantom considered two main parts: Image quality and fusion accuracy. Image quality part included for daily CT number linearity and spatial resolution, and fusion accuracy part designed to simulate a simple translation-rotation setting. The CT scans of the phantom obtained from conventional CT, MVCT of Tomotherapy unit, and both image sets were satisfied the recommendation of spatial resolution. This phantom was simple and efficient for daily imaging QA, and it is important to provide a new concept of verification of image registration.

SELECTION OF CITATIONS
SEARCH DETAIL
...