In order to compare the radiation immunity test level of ISO 14708 series standards and the domestic and international environmental radiation standards and then to ascertain whether the radiated immunity test level has reached the limit of the radiation strength in the relevant radiation environmental standards, this paper calculated the radiation field intensity and power density according to the radiated immunity test of ISO 14708 standards, and compared with the limit of ICNIRP 1998 and GB 8702-2014.


Subject(s)
Electromagnetic Radiation , Prostheses and Implants , Electromagnetic Fields , Immunity , Radiation Dosage
2.
Article in Chinese | WPRIM (Western Pacific) | ID: wpr-689841

ABSTRACT

In order to compare the radiation immunity test level of ISO 14708 series standards and the domestic and international environmental radiation standards and then to ascertain whether the radiated immunity test level has reached the limit of the radiation strength in the relevant radiation environmental standards, this paper calculated the radiation field intensity and power density according to the radiated immunity test of ISO 14708 standards, and compared with the limit of ICNIRP 1998 and GB 8702-2014.