Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Biomed Khim ; 60(3): 378-88, 2014.
Article in Russian | MEDLINE | ID: mdl-25019401

ABSTRACT

Methods of whole blood trace element determination by Graphite furnace atomic absorption spectrometry (in the variant of Zeeman's modulation polarization spectrometry) have been proposed. They do not require preliminary sample digestion. Furnace programs, modifiers and blood dilution factors were optimized. Seronorm™ human whole blood reference materials were used for validation. Dynamic ranges (for undiluted blood samples) were: Al 8 ¸ 210 мg/L; Be 0.3 ¸ 50 мg/L; Cd 0.2 ¸ 75 мg/L; Сo 5 ¸ 350 мg/L; Cr 10 ¸ 100 мg/L; Mn 6 ¸ 250 мg/L; Ni 10 ¸ 350 мg/L; Pb 3 ¸ 240 мg/L; Se 10 ¸ 500 мg/L; Tl 2 ¸ 600 мg/L. Precision (RSD) for the middle of dynamic range ranged from 5% for Mn to 11 for Se.


Subject(s)
Chalcogens/blood , Metals, Heavy/blood , Metals, Light/blood , Spectrophotometry, Atomic/methods , Humans , Reference Standards , Reference Values , Reproducibility of Results , Sensitivity and Specificity , Spectrophotometry, Atomic/instrumentation
SELECTION OF CITATIONS
SEARCH DETAIL
...