ABSTRACT
The comparison between the MEIA II and the EMIT assays for tacrolimus measurement and the interference by the hematocrit were evaluated in 93 samples from routine therapeutic monitoring at tacrolimus concentrations less than 9 microg/L (group A). Additionally, the incidence of false-positive results were determined in samples (n=46) from patients who were not receiving the drug (group B). In group A, no statistical differences were observed between the mean+/-SD values obtained by MEIA II (5.14+/-2.28 microg/L) and EMIT (4.61+/-1.79 microg/L). The correlation coefficient and the regression equation (95% CI) between both assays, were 0.761 and EMIT=1.088 (0.90, 1.35) MEIA II -0.38 (-1.65, -0.46), respectively. When the samples were stratified according to the hematocrit, the median differences between the methods (MEIA II minus EMIT) were as follows: hematocrit