1.
Medicine (Baltimore) ; 103(27): e38847, 2024 Jul 05.
Article in English | MEDLINE | ID: mdl-38968453

ABSTRACT

INTRODUCTION: The use of the PreserFlo microshunt is gaining popularity owing to its ease of implantation and reduced need for postoperative intervention compared to conventional trabeculectomy. PATIENT CONCERNS: However, microshunt exposure remains a severe complication of PreserFlo surgery, particularly in patients with a thin Tenon capsule and conjunctiva. However, the actual thickness and intensity of the Tenon capsule or conjunctiva can be confirmed only during surgery. DIAGNOSIS: Exfoliation glaucoma with previous several glaucoma surgeries with thinner Tenon capsule or conjunctiva. INTERVENTIONS: We performed PreserFlo implantation with a surgical technique to recover a thin Tenon capsule and conjunctiva by creating a half-thickness rectangular scleral flap under the shunt and covering it over the microshunt until the distal part, similar to the bridge. OUTCOMES: The patient had better intraocular pressure control with positive cosmetic appearance using this technique. CONCLUSION: This technique will be beneficial for both preventing exposure and holding down the top, in addition to improving cosmetic appearance.


Subject(s)
Sclera , Humans , Sclera/surgery , Trabeculectomy/methods , Intraocular Pressure , Exfoliation Syndrome/surgery , Glaucoma Drainage Implants/adverse effects , Tenon Capsule , Surgical Flaps , Postoperative Complications/prevention & control , Female , Aged , Male , Conjunctiva/surgery
3.
Int Ophthalmol ; 44(1): 283, 2024 Jun 26.
Article in English | MEDLINE | ID: mdl-38922523

ABSTRACT

PURPOSE: To examine the rate of ciliary body detachment in patients with choroidal detachment following glaucoma surgery and its effect on the clinical course, management, and prognosis. METHODS: A prospective observational case-series study. Patients with choroidal detachment following glaucoma surgery in 2018-2019 were included. All underwent complete ophthalmological examination and ultrasound biomicroscopy for evaluation of the presence and extent of ciliary body detachment. Follow-up examinations including ultrasound biomicroscopy scans were performed at 1 week, 1 month, 3 months, and 6 months. RESULTS: Eight patients (8 eyes) were enrolled, 4 male and 4 female, of mean age 72 years (range 60-83). Five patients underwent trabeculectomy with mitomycin C (0.02%), which was combined with phacoemulsification cataract extraction in one; two underwent Ahmed glaucoma valve implantations, and one underwent ab-interno Xen45 gel stent implantation with mitomycin C (0.02%). The mean intraocular pressure was 26.0 ± 7.65 mmHg preoperatively, dropping to 6.9 ± 2.64 mmHg on first postoperative day one. Mean time from surgery to diagnosis of choroidal detachment was 11.6 ± 5.73 days. Ciliary body detachment was identified by ultrasound biomicroscopy in all patients, ranging between one and four quadrants. All patients were treated with topical steroids and cycloplegics; three (37.5%) received oral steroids. No surgical intervention for the choroidal or ciliary body detachments was indicated. CONCLUSIONS: In this real-world prospective study, concurrent ciliary body detachment was identified in all patients who presented with choroidal detachment following glaucoma surgery. This observation may deepen our understanding of the mechanism underlying the hypotony that is often seen after glaucoma surgery.


Subject(s)
Choroidal Effusions , Ciliary Body , Glaucoma , Intraocular Pressure , Humans , Male , Female , Aged , Prospective Studies , Middle Aged , Aged, 80 and over , Intraocular Pressure/physiology , Choroidal Effusions/diagnosis , Choroidal Effusions/etiology , Glaucoma/surgery , Glaucoma/physiopathology , Glaucoma/complications , Postoperative Complications/diagnosis , Microscopy, Acoustic , Follow-Up Studies , Trabeculectomy/adverse effects , Trabeculectomy/methods , Glaucoma Drainage Implants/adverse effects , Visual Acuity , Uveal Diseases/diagnosis , Uveal Diseases/etiology , Tomography, Optical Coherence/methods
5.
Arq Bras Oftalmol ; 87(6): e2021, 2024.
Article in English | MEDLINE | ID: mdl-38597522

ABSTRACT

Implantation of glaucoma drainage devices is a valuable therapeutic option, particularly in children with glaucoma refractory to primary surgical treatment. Glaucoma drainage devices are typically used when conjunctival scarring hampers filtration surgery or prior angle procedures are not effective in controlling intraocular pressure. Despite known complications, the use of glaucoma drainage devices in children has increased in recent years, even as the primary surgical option. In this review, we evaluate the results of recent studies involving the implantation of glaucoma drainage devices in children, discussing new advances, and comparing the success rates and complications of different devices.


Subject(s)
Conjunctival Diseases , Glaucoma Drainage Implants , Glaucoma , Child , Humans , Glaucoma Drainage Implants/adverse effects , Glaucoma/surgery , Intraocular Pressure , Prosthesis Implantation/methods , Retrospective Studies , Treatment Outcome
6.
Arq Bras Oftalmol ; 87(3): e20230033, 2024.
Article in English | MEDLINE | ID: mdl-38537046

ABSTRACT

PURPOSE: This study aims to compare the initial ocular discomfort symptoms resulting from trabeculectomy and Ahmed glaucoma valve implantation surgeries. METHODS: A prospective comparative study was conducted. The evaluation of ocular discomfort employed a questionnaire designed to identify the frequency and severity of distinct symptoms: ocular pain, general discomfort, tearing, foreign body sensation, and burning. This questionnaire was administered prior to surgery as a baseline, and subsequently at 7, 30, and 90 days post-surgery. Simultaneously, the Ocular Surface Disease Index (OSDI) was applied at these same time intervals. RESULTS: The study encompassed a total of 17 patients (9 undergoing trabeculectomy and 8 undergoing Ahmed glaucoma valve implantation). The Ahmed glaucoma valve implantation group exhibited higher tearing levels at baseline (p=0.038). However, no statistically significant differences in symptoms were observed between the two surgeries at 7 and 30 days post-surgery. At the 90-day mark following surgery, patients who had undergone trabeculectomy reported a significantly higher foreign body sensation (p=0.004). Although OSDI scores did not differ between groups at baseline, the trabeculectomy group showed significantly higher OSDI scores than the Ahmed glaucoma valve implantation group at 7, 30, and 90 days after surgery (p<0.05). CONCLUSION: Post-surgery, patients who had undergone trabeculectomy experienced increased foreign body sensation. Trabeculectomy appears to cause greater early postoperative ocular discomfort compared to the Ahmed glaucoma valve implantation group.


Subject(s)
Foreign Bodies , Glaucoma Drainage Implants , Glaucoma , Trabeculectomy , Humans , Trabeculectomy/adverse effects , Prospective Studies , Intraocular Pressure , Glaucoma Drainage Implants/adverse effects , Glaucoma/etiology , Prosthesis Implantation/adverse effects , Foreign Bodies/complications , Foreign Bodies/surgery , Treatment Outcome , Retrospective Studies
7.
Medicine (Baltimore) ; 103(12): e37457, 2024 Mar 22.
Article in English | MEDLINE | ID: mdl-38518004

ABSTRACT

PURPOSE: Cosmetic iris implants have a record of high ocular complications and are no longer in use. These complications include glaucoma, corneal decompensation, iris atrophy, uveitis, cataract and retinal detachment. CASE PRESENTATION: We report a case of a 44-year-old lady presented with bilateral total iris atrophy, glaucoma and corneal decompensation after cosmetic artificial iris implantation. The patient underwent bilateral artificial iris removal, glaucoma drainage device for the right eye, and micropulse laser for the left eye. In addition, she underwent phacoemulsification with iris-diaphragm intraocular lens implant for the right. The cornea of the right eye ended up with successful Boston keratoprosthesis after rejection of previous 2 grafts. CONCLUSIONS: To the best of our knowledge, we describe the first report of bilateral total iris atrophy following a cosmetic iris implant accompanied by bilateral glaucoma and corneal decompensation.


Subject(s)
Corneal Diseases , Glaucoma Drainage Implants , Glaucoma , Iris Diseases , Female , Humans , Adult , Corneal Diseases/complications , Cornea , Device Removal/adverse effects , Intraocular Pressure , Glaucoma/etiology , Glaucoma/surgery , Iris/surgery , Glaucoma Drainage Implants/adverse effects
8.
Int Ophthalmol ; 44(1): 150, 2024 Mar 20.
Article in English | MEDLINE | ID: mdl-38503938

ABSTRACT

PURPOSE: To assess and compare the risk for development of cystoid macula edema (CME) after glaucoma drainage device (GDD) implantation versus conventional trabeculectomy with mitomycin (trab) for glaucoma. METHODS: Retrospective review of consecutive patients receiving trab or GDD implantation between 2016 and 2018. Inclusion criteria were availability of pre- and postoperative spectral domain optical coherence tomography (SD-OCT) of the macula. SD-OCT images were evaluated for presence of CME qualitatively, central subfield thickness (CST) and macular volume (MV). RESULTS: 73 eyes could be included, 42 received trab and 31 GDD surgery. Eyes receiving trab on average had 0.8 ± 0.8 previous intraocular operations, while eyes with GDD implantation had 3.1 ± 1.9 (p < 0.001). Occurrence of postoperative CME was significantly more frequent after GDD implantation (6 out of 31 (19.4%)) than after trab (2 out of 42 eyes = 4.8%), (p = 0.049). Mean preoperative CST as well as MV was comparable in both groups (CST before trab: 282.7 ± 23.0 µm, CST before GDD 284.2 ± 27.3 µm, p = 0.287; MV before trab: 7.8 ± 1.1 mm3, MV before GDD: 8.0 ± 0.8mm3, p = 0.305). Mean postoperative CST and MV were significantly higher after GDD (CST 338.5 ± 129.3 µm, MV 8.8 ± 2.6 mm3) than after trabeculectomy (CST 290.6 ± 60.2 µm, p = 0.038; MV 7.8 ± 1.2mm3, p = 0.039). CONCLUSIONS: In real-life conditions, GDD surgery seems to be associated with a higher risk to develop CME when compared to conventional trabeculectomy. This information may be helpful for glaucoma surgeons to advise the patients on postoperative risks of surgery.


Subject(s)
Glaucoma Drainage Implants , Glaucoma , Macular Edema , Trabeculectomy , Humans , Trabeculectomy/adverse effects , Trabeculectomy/methods , Intraocular Pressure , Glaucoma/surgery , Glaucoma Drainage Implants/adverse effects , Macular Edema/diagnosis , Macular Edema/etiology , Macular Edema/surgery , Edema , Retrospective Studies
9.
Acta Ophthalmol ; 102(5): e797-e804, 2024 Aug.
Article in English | MEDLINE | ID: mdl-38306110

ABSTRACT

PURPOSE: Bleb failure is a common complication after glaucoma filtration surgery. Different bleb classification schemes incorporating filtration bleb vascularization have been proposed, but the reported correlation with intraocular pressure (IOP) has been variable, possibly because of subjective vascularization grading. The purpose of the present study was to evaluate bleb vascularization after Preserflo Microshunt (PM) implantation using anterior segment OCT-angiography (AS-OCTA) as a biomarker for bleb failure. METHODS: Twenty-three eyes of twenty-three patients underwent PM implantation. Up to 12 months after surgery PM scleral passage-centred AS-OCTA measurements (PLEX Elite 9000) for bleb-vessel density (BVD) determination were performed and IOP as well as necessity for surgical revisions (needling and open revision) were documented. After multi-step image analysis (region of interest definition, artefact removal, binarization, BVD calculation), the predictive value of early postoperative BVD for surgical revisions was assessed using logistic regression modelling. RESULTS: Baseline IOP (23.57 ± 7.75 mmHg) decreased significantly to 8.30 ± 2.12, 9.17 ± 2.33 and 11.70 ± 4.40 mmHg after 1, 2 and 4 week(s), and 13.48 ± 5.83, 11.87 ± 4.49, 12.30 ± 6.65, 11.87 ± 3.11 and 13.05 ± 4.12 mmHg after 2, 3, 6, 9 and 12 month(s), respectively (p < 0.001). Nine patients (39%) needed surgical revisions after a median time of 2 months. Bleb vessel densities at 2 and 4 weeks were significantly associated with future surgical revisions upon logistic regression analysis (2 W/4 W likelihood-ratio test p-value: 0.0244/0.0098; 2 W/4 W area under the receiver operating characteristics curve: 0.796/0.909). CONCLUSION: Filtration bleb vessel density can be determined using AS-OCTA in the early postoperative period and is predictive for bleb failure after PM implantation.


Subject(s)
Intraocular Pressure , Reoperation , Tomography, Optical Coherence , Humans , Female , Male , Intraocular Pressure/physiology , Aged , Tomography, Optical Coherence/methods , Middle Aged , Fluorescein Angiography/methods , Follow-Up Studies , Glaucoma/surgery , Glaucoma/physiopathology , Glaucoma/diagnosis , Glaucoma Drainage Implants/adverse effects , Filtering Surgery/methods , Prospective Studies , Fundus Oculi , Conjunctiva/blood supply , Conjunctiva/surgery , Microvascular Density
10.
Sci Rep ; 14(1): 4452, 2024 02 23.
Article in English | MEDLINE | ID: mdl-38396005

ABSTRACT

To carry out the preclinical and histological evaluation of a novel nanotechnology-based microshunt for drainage glaucoma surgery. Twelve New Zealand White rabbits were implanted with a novel microshunt and followed up for 6 weeks. The new material composite consists of the silicone polydimethylsiloxane (PDMS) and tetrapodal Zinc Oxide (ZnO-T) nano-/microparticles. The microshunts were inserted ab externo to connect the subconjunctival space with the anterior chamber. Animals were euthanized after 2 and 6 weeks for histological evaluation. Ocular health and implant position were assessed at postoperative days 1, 3, 7 and twice a week thereafter by slit lamp biomicroscopy. Intraocular pressure (IOP) was measured using rebound tonometry. A good tolerability was observed in both short- and medium-term follow-up. Intraocular pressure was reduced following surgery but increased to preoperative levels after 2 weeks. No clinical or histological signs of inflammatory or toxic reactions were seen; the fibrotic encapsulation was barely noticeable after two weeks and very mild after six weeks. The new material composite PDMS/ZnO-T is well tolerated and the associated foreign body fibrotic reaction quite mild. The new microshunt reduces the IOP for 2 weeks. Further research will elucidate a tube-like shape to improve and prolong outflow performance and longer follow-up to exclude medium-term adverse effects.


Subject(s)
Glaucoma Drainage Implants , Glaucoma , Zinc Oxide , Animals , Rabbits , Glaucoma Drainage Implants/adverse effects , Glaucoma/surgery , Glaucoma/etiology , Intraocular Pressure , Tonometry, Ocular , Anterior Chamber/surgery , Nanotechnology
11.
Int Ophthalmol ; 44(1): 45, 2024 Feb 09.
Article in English | MEDLINE | ID: mdl-38336985

ABSTRACT

PURPOSE: To describe the in toto explantation of the CyPass® Micro-Stent and its conceivable complications. METHODS: This is a case series of eighteen eyes from fourteen patients who underwent CyPass® Micro-Stent implantation due to mild to moderate glaucoma and who subsequently suffered from loss of endothelial cell density. Consequently, the CyPass® Micro-Stent was in toto explanted. The surgical procedure and its complications are described and compared with trimming of the CyPass® Micro-Stent. RESULTS: A postoperative hyphema was developed in 8 of the 18 eyes. In four of them the hyphema was self-limiting, while in two patients an anterior chamber irrigation was necessary. One patient suffered from a severe intracameral bleeding and iridodialysis during explantation, so that the base of the iris had to be scleral fixated. The remaining explantations were without complications. CONCLUSION: Dealing with implanted CyPass® Micro-Stents poses a challenge for ophthalmic surgeons. An in toto removal can be traumatic, since the CyPass stent often is fibrotic encapsulated and fused with the surrounding tissue. Alternatively, trimming of the CyPass is also a viable option to avoid further endothelial damage. Reported complications of CyPass trimming are consistent with those that can occur after explantation. Further data on the development of the endothelial cells after trimming or explantation are not yet available. Therefore, it remains open whether trimming of the CyPass, in contrast to complete removal, carries the risk of further endothelial cell loss.


Subject(s)
Glaucoma Drainage Implants , Glaucoma, Open-Angle , Quinolinium Compounds , Thiazoles , Humans , Intraocular Pressure , Endothelial Cells , Hyphema , Glaucoma, Open-Angle/surgery , Glaucoma Drainage Implants/adverse effects , Anterior Chamber , Stents/adverse effects , Postoperative Complications
12.
Sci Rep ; 14(1): 3777, 2024 02 15.
Article in English | MEDLINE | ID: mdl-38355702

ABSTRACT

Glaucoma drainage devices (GDDs) are prosthetic-treatment devices for treating primary open-angle glaucoma. Despite their effectiveness in reducing intraocular pressures (IOP), endothelial cell damage (ECD) is a commonly known side-effect. There have been different hypotheses regarding the reasons for ECD with one being an induced increase in shear on the corneal wall. A computational fluid dynamics (CFD) model was used to investigate this hypothesis in silico. The Ahmed Glaucoma Valve (AGV) was selected as the subject of this study using an idealised 3D model of the anterior chamber with insertion angles and positions that are commonly used in clinical practice. It was found that a tube-cornea distance of 1.27 mm or greater does not result in a wall shear stress (WSS) above the limit where ECD could occur. Similarly, a tube-cornea angle of 45° or more was shown to be preferable. It was also found that the ECD region has an irregular shape, and the aqueous humour flow fluctuates at certain insertion angles and positions. This study shows that pathological amounts of WSS may occur as a result of certain GDD placements. Hence, it is imperative to consider the associated fluid force interactions when performing the GDD insertion procedure.


Subject(s)
Glaucoma Drainage Implants , Glaucoma, Open-Angle , Glaucoma , Humans , Glaucoma, Open-Angle/surgery , Hydrodynamics , Glaucoma Drainage Implants/adverse effects , Glaucoma/surgery , Intraocular Pressure , Endothelial Cells , Treatment Outcome , Retrospective Studies , Follow-Up Studies , Prosthesis Implantation
14.
Ophthalmol Glaucoma ; 7(1): 37-46, 2024.
Article in English | MEDLINE | ID: mdl-37567499

ABSTRACT

PURPOSE: To compare outcomes of early aqueous suppression (EAS) and standard therapy (ST) after Ahmed Glaucoma Valve (AGV) implantation for uveitic glaucoma. DESIGN: Retrospective comparative cohort study. PARTICIPANTS: All patients with uveitic glaucoma underwent AGV implantation from January 2010 to October 2020 at Northwestern Medicine. METHODS: Excluding the first postoperative day 1 (POD1), only eyes with IOP 10-15 mmHg at their first visit with IOP ≥ 10 mmHg were included in the main analysis. Early aqueous suppression (EAS) was defined as initiation of ocular hypotensive therapy when IOP was first 10-15 mmHg. Standard therapy was initiation of therapy at any later time. Failure was defined as IOP > 21 mmHg, < 5 mmHg, or < 20% reduction in IOP from baseline after 3 months, for 2 consecutive study visits. Hypotony was defined as IOP ≤ 5 mmHg for ≥ 2 visits. Hypertensive phase was defined as IOP > 21 mmHg for 2 consecutive visits in the first 3 months. MAIN OUTCOME MEASURES: Proportion achieving overall success; incidence of hypotony and hypertensive phase. RESULTS: Twenty-eight eyes of 26 patients were in the EAS group and 20 eyes of 19 patients were in the ST group, with a mean follow-up of 17.7 and 28.2 months, respectively. Baseline IOP was similar in the EAS (31.2 ± 10.1 mmHg) and ST (34.6 ± 12.2 mmHg) groups; P = 0.18. Final IOP was lower in the EAS group (12.9 ± 4.6 mmHg) than the ST group (16.4 ± 5.7 mmHg; P = 0.02) on 2.6 ± 0.9 medications in the EAS group and 1.8 ± 1.5 in the ST group (P = 0.07). Overall success was achieved in 87% of EAS eyes and 74% of ST eyes (P = 0.43). There were no statistically significant differences in the occurrence of additional glaucoma surgery (4% for EAS, 20% for ST; P = 0.11), hypotony (7% for EAS, 0% for ST; P = 0.50), or hypertensive phase (4% for EAS, 21% for ST; P = 0.09). CONCLUSIONS: EAS was associated with a lower final IOP after AGV in uveitic glaucoma eyes; however, more medications were in use at the final visit. No statistically significant differences in overall success or the incidence of adverse events were observed. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.


Subject(s)
Glaucoma Drainage Implants , Glaucoma , Humans , Retrospective Studies , Cohort Studies , Intraocular Pressure , Glaucoma/etiology , Glaucoma/surgery , Glaucoma Drainage Implants/adverse effects
15.
Eur J Ophthalmol ; 34(1): NP27-NP31, 2024 Jan.
Article in English | MEDLINE | ID: mdl-37192674

ABSTRACT

BACKGROUND: To discuss the efficacy of an inferior implant of XEN 63 gel stent in a patient with refractory glaucoma after trabeculectomy failure and vitreoretinal surgery with silicone oil tamponade. CASE DESCRIPTION: We report the case of a 73-year-old man with a history of refractory open glaucoma with trabeculectomy failure. He experienced recurrent retinal detachments with silicone oil tamponade, with uncontrolled intra-ocular pressure (IOP) after silicone oil removal. Due to the presence of oil emulsion in the anterior chamber, the chosen location for XEN 63 implantation was the infero-temporal quadrant. Mild hyphema and vitreous hemorrhage were seen post-operatively, but were self-limiting. At week 1, the intraocular pressure was 8 mmHg with a well-formed bleb seen in anterior segment optical coherence tomography (AS-OCT). At 6 month follow up, the patient maintained a IOP of 12 mmHg without topical hypotensive drugs. Slit lamp examination revealed a widespread, developed bleb with no signs of inflammation. CONCLUSION: In this case of refractory glaucoma in a vitrectomized eye with previous oil tamponade, the inferior placement of the XEN 63 gel stent delivered an adequate intraocular pressure even at 6-months follow up, with a diffuse functional infero-nasal bleb seen with AS-OCT.


Subject(s)
Glaucoma Drainage Implants , Glaucoma, Open-Angle , Glaucoma , Male , Humans , Aged , Glaucoma, Open-Angle/surgery , Silicone Oils/adverse effects , Glaucoma Drainage Implants/adverse effects , Glaucoma/surgery , Intraocular Pressure , Stents/adverse effects
16.
J Glaucoma ; 33(2): 132-138, 2024 02 01.
Article in English | MEDLINE | ID: mdl-37974333

ABSTRACT

PRCIS: In this study, we report a 2-year follow-up after implantation of an adjustable glaucoma drainage device combined with a novel orbital filtering plate. The IOP was efficiently controlled postoperatively with limited complications and an excellent safety profile. PURPOSE: To report the clinical results at 2 years after implantation of an eyeWatch (eW) system, which comprises an adjustable glaucoma drainage device (eW implant) combined with a novel orbital filtering plate (eyePlate), in refractory glaucoma. PATIENTS AND METHODS: Monocentric, prospective, noncomparative clinical trial. Patients suffering from refractory glaucoma after several failed surgeries were operated using the eW system. The primary outcome was the success rate, defined as an IOP ≤18 mm Hg and reduction of more than 20% from baseline, IOP ≥6 mm Hg. Secondary outcomes were mean IOP, visual acuity, the number of antiglaucoma medications, and the number and type of complications. RESULTS: Forty-two eyes from 42 patients were included. The mean follow-up time was 22.0±4.0 months. The mean baseline IOP decreased from 23.2±7.2 mm Hg before surgery to 11.9±4.4 mm Hg at 12 months ( P <0.001) and 11.5±3.0 at 24 months. The mean number of glaucoma medications decreased from 2.9±0.9 before surgery to 0.7±0.9 at 12 months ( P <0.001) and 1.0±0.9 at 24 months. The qualified success rate was 93% at 12 months and 90% at 24 months whereas the complete success rate was 46% and 40% at 12 and 24 months, respectively. Complications occurred in 7 patients. CONCLUSIONS: Implantation of the eW system effectively lowered IOP and reduced glaucoma medications for a 2-year period in patients who had previously failed glaucoma surgery. Persistent hypotony was not observed and the number of complications was low.


Subject(s)
Glaucoma Drainage Implants , Glaucoma , Humans , Prospective Studies , Intraocular Pressure , Prosthesis Implantation/methods , Treatment Outcome , Postoperative Complications/etiology , Glaucoma/complications , Glaucoma Drainage Implants/adverse effects , Follow-Up Studies , Retrospective Studies
17.
Indian J Ophthalmol ; 72(3): 370-373, 2024 Mar 01.
Article in English | MEDLINE | ID: mdl-38099583

ABSTRACT

BACKGROUND: Keiki Mehta body pressure (BP) valve is a glaucoma drainage device that works on the simple mechanism of body pressure. PURPOSE: To evaluate the short- to intermediate-term efficacy and safety of Keiki Mehta valve implantation in treating refractory glaucoma. METHODS: This was a prospective, interventional, institution-based study. Adult patients (age >18 years) having refractory glaucoma who underwent Keiki Mehta valve implantation from May 2019 to December 2019 in a tertiary hospital in eastern India were followed up to 18 months. RESULTS: Twenty patients (20 eyes) with refractory glaucoma who underwent Keiki Mehta valve implantation were followed up on postoperative day 1 and at 1, 3, 6, 12, and 18 months postoperatively. Visual acuity, intraocular pressure (IOP), antiglaucoma medications, and success rate were measured at every visit. There was significant reduction in mean IOP from the preoperative values at every postoperative follow-up ( P < 0.05). Kaplan-Meier survival analysis demonstrated a probability of 50% complete success and 20% qualified success at 18 months. Hypotony was the most common postoperative complication. CONCLUSION: Keiki Mehta valve implantation is an effective and safe treatment for refractory glaucoma.


Subject(s)
Glaucoma Drainage Implants , Glaucoma , Adult , Humans , Adolescent , Prospective Studies , Treatment Outcome , Follow-Up Studies , Prosthesis Implantation , Retrospective Studies , Glaucoma/etiology , Glaucoma Drainage Implants/adverse effects , Intraocular Pressure , Postoperative Complications/surgery
18.
BMC Ophthalmol ; 23(1): 509, 2023 Dec 14.
Article in English | MEDLINE | ID: mdl-38097974

ABSTRACT

BACKGROUND: To investigate the outcome of non-valved glaucoma drainage implant surgery (GDIS) in primary open-angle glaucoma (POAG) patients divided in the first GDI group (patients who underwent the first GDIS) and the second GDI group (patients who underwent the second GDIS because of the failed first GDIS). METHODS: Intraocular pressure (IOP), visual acuity (VA), visual field defect (VFD), medication score (MS), survival rate of GDIS, complications, and patient background was retrospectively analyzed. Two success criteria were set: Criteria (1) IOP reduction ≥ 20% and 5 < IOP ≤ 21, Criteria (2) IOP reduction ≥ 20% and 5 < IOP ≤ 14. RESULTS: There were 136 eyes of 109 patients in the first GDI group and 32 eyes of 27 patients in the second GDI group. In the first GDI group and II, mean preoperative IOP was 26.7 ± 6.7 mmHg and 23.7 ± 3.5 mmHg, respectively (P = 0.09). No statistically significant difference in postoperative IOP reduction was found between the two groups (P = 0.39). At 5-years postoperative, the Criteria 1 (Criteria 2) survival rate in the first GDI group and the second GDI group was 60.4% (31.7%) and 61.2% (25.6%), respectively (Criteria 1: hazard ratio [HR]: 0.64, 95% confidence interval [CI]: 0.30-1.35 [P = 0.24]; Criteria 2: HR: 0.81, 95% CI: 0.46-1.44, P = 0.48). No significant difference in VA, VFD change, MS, or complications was observed. Young patient age was the only significant factor for failure in the first GDI group (odds ratio: 0.95, 95% confidence interval: 0.91-1.00, P = 0.03). CONCLUSION: The second GDIS may be as effective as the first GDIS for IOP reduction in POAG patients, however, there is a high risk of failure in young-age patients and the surgery may be ineffective in eyes requiring Criteria 2.


Subject(s)
Glaucoma Drainage Implants , Glaucoma, Open-Angle , Glaucoma , Ocular Hypotension , Trabeculectomy , Humans , Glaucoma/surgery , Glaucoma, Open-Angle/surgery , Glaucoma, Open-Angle/etiology , Retrospective Studies , Glaucoma Drainage Implants/adverse effects , Intraocular Pressure , Ocular Hypotension/etiology , Guanine Nucleotide Dissociation Inhibitors , Treatment Outcome
19.
BMC Ophthalmol ; 23(1): 465, 2023 Nov 16.
Article in English | MEDLINE | ID: mdl-37974114

ABSTRACT

BACKGROUND: Choroidal effusion is a common complication of glaucoma surgery. Although most cases of choroidal effusions resolve spontaneously with observation or medical management alone as intraocular pressure normalizes, surgical drainage might be needed in severe or persistent cases. Herein, we report a case of spontaneous resolution of long-standing severe choroidal effusion after Ahmed glaucoma valve implantation. CASE PRESENTATION: An 85-year-old man with uncontrolled primary open-angle glaucoma and medical history of chronic kidney disease underwent uneventful Ahmed glaucoma valve implantation. On postoperative day 8, transient hypotony occurred, and large 360° peripheral choroidal detachments developed. Although the intraocular pressure increased to normal levels on postoperative day 15, choroidal effusion did not resolve. Fundus examination over 8 months showed that the large choroidal effusion persisted despite a well-controlled intraocular pressure. Laboratory test performed at preoperatively and follow-up period revealed persistently elevated potassium and creatinine levels. On postoperative 9 months, the lesion resolved spontaneously without any surgical intervention. We found that the patient's creatinine level was normalized, pre-existing hyperkalemia was corrected, and accordingly his general condition was improved. CONCLUSIONS: Considering the underlying medical condition may be helpful in patients with persistent choroidal effusion of an unclear etiology following glaucoma filtering surgery.


Subject(s)
Choroidal Effusions , Glaucoma Drainage Implants , Glaucoma, Open-Angle , Glaucoma , Male , Humans , Aged, 80 and over , Glaucoma, Open-Angle/surgery , Creatinine , Postoperative Complications , Glaucoma Drainage Implants/adverse effects , Glaucoma/surgery , Intraocular Pressure , Choroidal Effusions/diagnosis , Choroidal Effusions/etiology , Drainage , Treatment Outcome , Retrospective Studies
20.
Vestn Oftalmol ; 139(5): 90-95, 2023.
Article in Russian | MEDLINE | ID: mdl-37942602

ABSTRACT

Glaucoma is a severe, rapidly progressing disease that in the absence of proper treatment leads to blindness in 20% of patients. According to the World Glaucoma Association, this disease is the most socially significant in modern ophthalmology and requires searching for new and effective methods of treatment. This article presents the results of research and reviews on this issue, considers both conservative therapy and surgical methods of treatment, analyzes in detail modern methods of micro-invasive eye surgery actively used in clinical practice. The article also describes indications for a various types of interventions, as well as the effect achieved by them and the possible complications, and presents the conclusions about the possibility of using these procedures in wide clinical practice.


Subject(s)
Glaucoma Drainage Implants , Glaucoma , Ophthalmology , Humans , Intraocular Pressure , Glaucoma Drainage Implants/adverse effects , Glaucoma/diagnosis , Glaucoma/surgery , Ophthalmologic Surgical Procedures/adverse effects , Stents