1.
Talanta ; 68(2): 480-7, 2005 Dec 15.
Artículo en Inglés | MEDLINE | ID: mdl-18970347

RESUMEN

In this work, we have modified the technique of all injection analysis (AIA) by changing the position of the detector. The detection is then located as a part of the circulatory loop. With this new detector position, we could monitor for many numbers of circulation. The sensitivity was improved by using the cumulative signal data obtained when the number of circulation rounds was increased. The dilution effect using this new detector location was also less than that with the previous system. We employed a four-channel peristaltic pump to aspirate four types of liquids into the system together at one time. The AIA method was then developed for determination of phosphorus in soils and sediment extracts. The method was optimized for the new harmonized scheme of extraction that has been developed by the European Commission.