1.
Ann Thorac Surg ; 32(6): 604-8, 1981 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-7316594

RESUMEN

We present the case of a patient who had rupture of a pulsatile assist device (PAD) accompanied by massive air embolism, and the treatment that brought it to a successful outcome. After rupture of the skin of the PAD balloon, a massive amount of air was injected into the ascending aorta. The patient was placed in Trendelenburg position and cooled in deep hypothermia with cardiopulmonary bypass. He was given 1 gm of methylprednisolone intravenously, and the aortic valve replacement and double vein bypass graft were performed. After completion of the operation, the patient was partially rewarmed to 30 degrees C central temperature and transported by ambulance to a hyperbaric chamber where he was compressed to 6 atmospheres absolute 9 hours after the accident with clinical signs of severe brain dysfunction. The patient recovered completely and was discharged from the hospital on the tenth postoperative day.


Asunto(s)
Circulación Asistida/efectos adversos , Embolia Aérea/terapia , Oxigenoterapia Hiperbárica , Oxigenadores/efectos adversos , Embolia Aérea/etiología , Humanos , Hipotermia Inducida , Complicaciones Intraoperatorias , Masculino , Metilprednisolona/uso terapéutico , Persona de Mediana Edad , Postura