1.
Ther Drug Monit ; 20(2): 191-201, 1998 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-9558134

RESUMEN

A novel fluorescence polarization immunoassay for vancomycin on Abbott AxSYM analyzer is described. The immunoassay allows for the accurate quantification of vancomycin in the presence of the crystalline degradation product (CDP). It displays dilution linearity from 1.0 microg/ml to 100.0 microg/ml, coefficients of variation ranging from 2.94% to 4.26%, recovery from 98% to 105%, and a sensitivity of <2.0 microg/ml. The assay demonstrates no cross-reactivity to crystalline degradation product, and to commonly-prescribed and over-the-counter drugs, as well as a minimum interference from endogenous substances.


Asunto(s)
Antibacterianos/sangre , Monitoreo de Drogas/métodos , Inmunoensayo de Polarización Fluorescente/métodos , Vancomicina/sangre , Animales , Antibacterianos/química , Calibración , Cromatografía Líquida de Alta Presión , Intervalos de Confianza , Cristalización , Contaminación de Medicamentos/prevención & control , Interacciones Farmacológicas , Monitoreo de Drogas/instrumentación , Femenino , Inmunoensayo de Polarización Fluorescente/instrumentación , Humanos , Modelos Lineales , Ratones , Medicamentos sin Prescripción/química , Reproducibilidad de los Resultados , Sensibilidad y Especificidad , Espectrofotometría Ultravioleta , Vancomicina/química