1.
Skin Res Technol ; 14(4): 385-9, 2008 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-18937771

RESUMEN

BACKGROUND/PURPOSE: The mechanical properties of the fingertip skin are very important when studying dexterous manipulation. These properties are strongly influenced by the level of skin hydration. Currently, there is no device capable of measuring skin moisture during object manipulation. METHODS: Skin moisture levels during object manipulation were measured using the Moisture Evaluator, a probe consisting of gold-covered electrodes connected to a resistor-capacitor circuit. In vivo calibration was performed by comparison with measurements obtained using a Corneometer at two normal force levels (0.2 and 2 N). RESULTS: Measurements from the Moisture Evaluator were well correlated with those from the Corneometer. CONCLUSION: A new device for evaluating skin moisture at the fingertip has been designed and validated.


Asunto(s)
Agua Corporal/metabolismo , Electrodiagnóstico/instrumentación , Dedos/fisiología , Pletismografía de Impedancia/instrumentación , Piel/metabolismo , Agua/análisis , Adulto , Electrodiagnóstico/métodos , Diseño de Equipo , Análisis de Falla de Equipo , Femenino , Humanos , Masculino , Pletismografía de Impedancia/métodos