J Synchrotron Radiat ; 21(Pt 2): 452-5, 2014 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-24562569

RESUMEN

A new bending-magnet beamline with a 2.5 m normal-incidence monochromator has been constructed to serve with a light source in the visible-vacuum-ultraviolet region for photoluminescence, transmission and reflection spectroscopies of solids at the UVSOR-III 750 MeV synchrotron radiation light source. The aim is to pave the way to establishing a beamline with high photon flux, high brilliance, high energy-resolution, high linear-polarization and low higher-order light. To obtain high photon flux and brilliance, the acceptance angle of the bending-magnet radiation was designed to be 40 mrad (H) × 14 mrad (V) and the post-mirror system employed Kirkpatrick-Baez optics. The incidence angle of the incoming light to the optical elements, except to the gratings, was set to a grazing angle in order to keep a degree of linear polarization. For achieving high energy-resolution, an off-plane Eagle-type monochromator was adopted. Higher-order unwanted light in the energy range below ∼11 eV was suppressed to be less than 0.1%.


Asunto(s)
Análisis Espectral/instrumentación , Sincrotrones , Diseño de Equipo , Magnetismo , Óptica y Fotónica , Fotones , Análisis Espectral/métodos , Rayos Ultravioleta