Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
J Chromatogr A ; 1710: 464405, 2023 Nov 08.
Artículo en Inglés | MEDLINE | ID: mdl-37769426

RESUMEN

Polysorbate 80 is widely used as a formulation component in biopharmaceutical drug products. Recent studies have shown that polysorbate 80 is readily degraded either through direct or indirect means. The degradation of polysorbate 80 causes a concern for the long-term stability of biopharmaceutical drug product, as the breakdown products of polysorbate 80 have been shown to cause adverse effects, such as formation of sub-visible and visible particles and mAb aggregation. Understanding the path and extent of degradation is of a paramount importance for the formulator during formulation development. A multi-detector HPLC system using charged aerosol and mass detection was developed and optimized for the characterization of polysorbate 80 standards. The system included a post-column make-up flow, i.e. an inverse gradient, that enabled constant eluent composition at the detectors. The inverse gradient eliminated the main source of variability for the charged aerosol detector response, thereby enabling the calculation of the mass balance between polysorbate components with different degrees of esterification. Extracted ion chromatograms of the mass detector combined with their respective retention times were used to qualitatively characterize the polysorbate samples down to the individual components. The system was applied to study the degradation of several polysorbate standards which occurred by enzymatic digestion or long-term storage. The system provided detailed information on the mechanism of degradation without the need for additional orthogonal analytical techniques.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA