1.
Rev Sci Instrum ; 85(6): 066107, 2014 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-24985871

RESUMEN

A conventional Fourier transform-Ion Cyclotron Resonance (ICR) detection cell is azimuthally divided into four equal sections. One pair of opposed electrodes is used for ion cyclotron excitation, and the other pair for ion image charge detection. In this work, we demonstrate that an appropriate electrical circuit facilitates excitation and detection on one pair of opposed electrodes. The new scheme can be used to minimize the number of electrically independent ICR cell electrodes and/or improve the electrode geometry for simultaneously increased ICR signal magnitude and optimal post-excitation radius, which results in higher signal-to-noise ratio and decreased space-charge effects.


Asunto(s)
Ciclotrones , Análisis de Fourier , Espectrometría de Masas , Electrodos , Espectrometría de Masas/instrumentación , Espectrometría de Masas/métodos