Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
Med Phys ; 46(11): 4725-4737, 2019 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-31446633

RESUMEN

PURPOSE: Kilovoltage intrafraction monitoring (KIM) allows for real-time image guidance for tracking tumor motion in six-degrees-of-freedom (6DoF) on a standard linear accelerator. This study assessed the geometric accuracy and precision of KIM used to guide patient treatments in the TROG 15.01 multi-institutional Stereotactic Prostate Ablative Radiotherapy with KIM trial and investigated factors affecting accuracy and precision. METHODS: Fractions from 44 patients with prostate cancer treated using KIM-guided SBRT were analyzed across four institutions, on two different linear accelerator models and two different beam models (6 MV and 10 MV FFF). The geometric accuracy and precision of KIM was assessed from over 33 000 images (translation) and over 9000 images (rotation) by comparing the real-time measured motion to retrospective kV/MV triangulation. Factors potentially affecting accuracy, including contrast-to-noise ratio (CNR) of kV images and incorrect marker segmentation, were also investigated. RESULTS: The geometric accuracy and precision did not depend on treatment institution, beam model or motion magnitude, but was correlated with gantry angle. The centroid geometric accuracy and precision of the KIM system for SABR prostate treatments was 0.0 ± 0.5, 0.0 ± 0.4 and 0.1 ± 0.3 mm for translation, and -0.1 ± 0.6°, -0.1 ± 1.4° and -0.1 ± 1.0° for rotation in the AP, LR and SI directions respectively. Centroid geometric error exceeded 2 mm for 0.05% of this dataset. No significant relationship was found between large geometric error and CNR or marker segmentation correlation. CONCLUSIONS: This study demonstrated the ability of KIM to locate the prostate with accuracy below other uncertainties in radiotherapy treatments, and the feasibility for KIM to be implemented across multiple institutions.


Asunto(s)
Fraccionamiento de la Dosis de Radiación , Neoplasias de la Próstata/fisiopatología , Neoplasias de la Próstata/radioterapia , Radiocirugia/métodos , Radioterapia Guiada por Imagen/métodos , Humanos , Masculino , Aceleradores de Partículas , Radiocirugia/instrumentación , Planificación de la Radioterapia Asistida por Computador , Radioterapia Guiada por Imagen/instrumentación , Estudios Retrospectivos
2.
Med Phys ; 45(3): 1222-1232, 2018 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-29363760

RESUMEN

PURPOSE: Most modern radiotherapy machines are built with a 2D kV imaging system. Combining this imaging system with a 2D-3D inference method would allow for a ready-made option for real-time 3D tumor tracking. This work investigates and compares the accuracy of four existing 2D-3D inference methods using both motion traces inferred from external surrogates and measured internally from implanted beacons. METHOD: Tumor motion data from 160 fractions (46 thoracic/abdominal patients) of Synchrony traces (inferred traces), and 28 fractions (7 lung patients) of Calypso traces (internal traces) from the LIGHT SABR trial (NCT02514512) were used in this study. The motion traces were used as the ground truth. The ground truth trajectories were used in silico to generate 2D positions projected on the kV detector. These 2D traces were then passed to the 2D-3D inference methods: interdimensional correlation, Gaussian probability density function (PDF), arbitrary-shape PDF, and the Kalman filter. The inferred 3D positions were compared with the ground truth to determine tracking errors. The relationships between tracking error and motion magnitude, interdimensional correlation, and breathing periodicity index (BPI) were also investigated. RESULTS: Larger tracking errors were observed from the Calypso traces, with RMS and 95th percentile 3D errors of 0.84-1.25 mm and 1.72-2.64 mm, compared to 0.45-0.68 mm and 0.74-1.13 mm from the Synchrony traces. The Gaussian PDF method was found to be the most accurate, followed by the Kalman filter, the interdimensional correlation method, and the arbitrary-shape PDF method. Tracking error was found to strongly and positively correlate with motion magnitude for both the Synchrony and Calypso traces and for all four methods. Interdimensional correlation and BPI were found to negatively correlate with tracking error only for the Synchrony traces. The Synchrony traces exhibited higher interdimensional correlation than the Calypso traces especially in the anterior-posterior direction. CONCLUSION: Inferred traces often exhibit higher interdimensional correlation, which are not true representation of thoracic/abdominal motion and may underestimate kV-based tracking errors. The use of internal traces acquired from systems such as Calypso is advised for future kV-based tracking studies. The Gaussian PDF method is the most accurate 2D-3D inference method for tracking thoracic/abdominal targets. Motion magnitude has significant impact on 2D-3D inference error, and should be considered when estimating kV-based tracking error.


Asunto(s)
Radioterapia/métodos , Ensayos Clínicos como Asunto , Humanos , Errores Médicos , Neoplasias/radioterapia , Distribución Normal , Factores de Tiempo , Rayos X
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA