1.
A A Pract ; 15(2): e01389, 2021 Feb 05.
Artículo en Inglés | MEDLINE | ID: mdl-33560644

RESUMEN

A patient received closed-circuit anesthesia from a General Electric Avance S/5 (GE Healthcare, Madison, WI) anesthesia machine during a robotic abdominal procedure. With return of spontaneous ventilation at the end of the procedure, the negative airway pressure alarm began to sound, and a negative airway pressure of 10-15 cm H2O was observed with each breath. Replacing the CO2 absorber resolved the problem. There was considerable condensation on the walls of the Amsorb canister, and on disassembly, the sponge at the bottom was wet. Experimentation with an empty canister revealed that as little as 30 mL of water in the sponge can reproduce our observations.


Asunto(s)
Hidróxido de Calcio , Agua , Cloruro de Calcio , Diseño de Equipo , Humanos