Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
Br J Ophthalmol ; 93(4): 502-5, 2009 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-19074920

RESUMEN

AIMS: The aim of the study was to compare the accuracy of A-scan biometry and MRI for the measurement of axial length in silicone oil-filled eyes. METHODS: This was a prospective randomised study of 70 patients. Biometry was performed using MRI in 33 patients (MRI group) and A-scan echography in 37 patients (A-scan group). The difference between predicted and final refraction was measured and evaluated statistically. RESULTS: In patients with axial length >/=26 mm, the mean deviation of the final from predicted refraction was -1.23 (SD 0.67) D in the MRI group and -2.3 (SD 2.02) D in the A-scan group. The difference between these two groups was statistically significant (p = 0.02). In patients with axial length <26 mm, the mean deviation of the final from predicted refraction was -0.12 (SD 1.29) D in the MRI group and -0.33 (SD 1.39) D in the A-scan group. There was no statistical significance between the two groups (p = 0.629). CONCLUSION: For highly myopic patients MRI biometry was a more accurate measurement of axial length in silicone oil-filled eyes. A-scan and MRI biometry were comparably accurate in measuring axial length in patients with axial length <26 mm.


Asunto(s)
Ojo/diagnóstico por imagen , Ojo/patología , Errores de Refracción/diagnóstico , Aceites de Silicona , Anciano , Biometría/métodos , Femenino , Humanos , Imagen por Resonancia Magnética/métodos , Masculino , Persona de Mediana Edad , Facoemulsificación , Estudios Prospectivos , Errores de Refracción/diagnóstico por imagen , Curvatura de la Esclerótica , Ultrasonografía , Vitrectomía
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA