RESUMEN
Metal injection molding (MIM) has become an important manufacturing technology for biodegradable medical devices. As a biodegradable metal, pure iron is a promising biomaterial due to its mechanical properties and biocompatibility. In light of this, we performed the first study that manufactured and evaluated the in vitro and in vivo biocompatibility of samples of iron porous implants produced by MIM with a new eco-friendly feedstock from natural rubber (Hevea brasiliensis), a promisor binder that provides elastic property in the green parts. The iron samples were submitted to tests to determine density, microhardness, hardness, yield strength, and stretching. The biocompatibility of the samples was studied in vitro with adipose-derived mesenchymal stromal cells (ADSCs) and erythrocytes, and in vivo on a preclinical model with Wistar rats, testing the iron samples after subcutaneous implant. Results showed that the manufactured samples have adequate physical, and mechanical characteristics to biomedical devices and they are cytocompatible with ADSCs, hemocompatible and biocompatible with Wistars rats. Therefore, pure iron produced by MIM can be considered a promising material for biomedical applications.