Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros










Base de datos
Intervalo de año de publicación
1.
Am J Psychiatry ; 157(6): 963-7, 2000 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-10831477

RESUMEN

OBJECTIVE: The maximum output charge for ECT devices is limited to 576 millicoulombs in the United States, although there are no data ensuring that this limit will allow consistently effective treatments. The authors examined whether this limit has a negative impact on therapeutic response and, therefore, whether a higher stimulus charge should be available. METHOD: They retrospectively reviewed the records of 471 patients who received a clinical index course of ECT at Duke University between 1991 and 1998. These patients received conservative stimulus dosing of 2.25 times seizure threshold for unilateral ECT and 1.5 times seizure threshold for bilateral ECT. RESULTS: Seventy-two (15%) of the 471 patients required the maximum stimulus intensity during their index ECT course. Of these, 24 (5% of the total) had either a short EEG seizure (less than 25 seconds) or had no seizure at the maximum level. Strategies to augment therapeutic response with caffeine, ketamine, or hyperventilation were used in 14 of the 24 patients, and data on therapeutic response were available for 22 of the 24. Only seven (32%) of these 22 patients were considered ECT responders, compared with 242 (66%) of the remaining 364 patients for whom data on response to ECT were available. Older age and pre-ECT course EEG slowing were predictors of requiring the maximum stimulus level. CONCLUSIONS: The maximum available stimulus output was therapeutically insufficient for 5% of the patients studied even when available means to augment response were instituted. This percentage would likely be even larger with the use of a less conservative dosing protocol for unilateral ECT. Increases in maximum stimulus output for ECT devices should be considered as a means to ensure adequate treatment response.


Asunto(s)
Terapia Electroconvulsiva/métodos , Terapia Electroconvulsiva/estadística & datos numéricos , Trastornos Mentales/terapia , Encéfalo/fisiología , Trastorno Depresivo/psicología , Trastorno Depresivo/terapia , Estimulación Eléctrica , Terapia Electroconvulsiva/instrumentación , Electroencefalografía/estadística & datos numéricos , Femenino , Lateralidad Funcional/fisiología , Humanos , Masculino , Trastornos Mentales/psicología , Persona de Mediana Edad , Análisis Multivariante , Análisis de Regresión , Estudios Retrospectivos , Esquizofrenia/terapia , Psicología del Esquizofrénico , Resultado del Tratamiento
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...