1.
J Vasc Interv Radiol ; 8(6): 1029-37, 1997.
Artículo en Inglés | MEDLINE | ID: mdl-9399474

RESUMEN

PURPOSE: A comparison of tilting, caval coverage, asymmetry, and insertion problems with the over-the-wire stainless-steel and titanium versions of the Greenfield filter. MATERIALS AND METHODS: The study compared 104 stainless-steel and 141 titanium Greenfield inferior vena cava (IVC) filter insertions. The angle the sheath and deployed filter made relative to the cava, as well as filter strut distribution, were determined from spot films. The proportionate caval coverage was computed from the cavogram (anteroposterior projection). Mean filter tilts, subgrouped by insertion site, and caval coverage were compared with the Student t test, whereas strut patterns were analyzed with a contingency table. RESULTS: The filter caval and sheath caval angles correlated. The filter caval angles varied with insertion site, but were lowest with a right jugular approach. Caval coverage was identical with both designs. The stainless-steel version resulted in a more uniform distribution of struts in comparison with the titanium version. The incidence of insertion problems was not significantly different between the filter types. CONCLUSIONS: While IVC filter tilting was not improved with the newer design, the pattern of struts was more uniformly symmetric with the stainless-steel device. The right jugular insertion site was associated with the lowest filter caval angles and the most symmetric pattern of struts.


Asunto(s)
Filtros de Vena Cava , Vena Cava Inferior , Distribución de Chi-Cuadrado , Femenino , Humanos , Masculino , Persona de Mediana Edad , Radiografía Intervencional , Estudios Retrospectivos , Acero Inoxidable , Titanio
4.
Invest Radiol ; 26(10): 874-8, 1991 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-1960028

RESUMEN

The limitations of angiography, when it is used as the sole method of vascular assessment, are increasingly apparent as vascular intervention becomes more sophisticated. Angioscopy could be an adjunctive diagnostic modality by differentiating among thrombus, dissection, and atheroma, and by monitoring the response to therapy. However, angioscopy requires a blood-free field for adequate visibility, and this may be difficult to obtain. The feasibility of carbon dioxide (CO2) and various saline delivery methods for clarifying the viewing field for percutaneous angioscopy was investigated. Angioscopy of femoral and iliac arteries on nine dogs was performed. Saline was infused by hand injection, pressure bag infusion, or mechanical power injection, and CO2 gas was injected using a special gas injector. The clarity of the viewing field was graded for each medium and method. Excellent quality antegrade femoral angioscopy was obtained with CO2. The superiority of CO2 injection in comparison with power-injected saline approached statistical significance (P = .06). Power-injected CO2 and power saline were superior to hand-injected or pressure bag-injected saline for maintaining sufficient visibility. Retrograde iliac angioscopy was possible without inflow occlusion, but required high flow rates (only possible with power-injected saline or CO2). CO2 injected under controlled circumstances holds promise as a medium to improve angioscopic visibility.


Asunto(s)
Vasos Sanguíneos , Dióxido de Carbono/administración & dosificación , Endoscopía , Angiografía , Animales , Perros , Endoscopía/métodos , Femenino , Arteria Femoral , Masculino , Cloruro de Sodio/administración & dosificación
5.
Toxicology ; 62(1): 89-94, 1990 May 14.
Artículo en Inglés | MEDLINE | ID: mdl-2343459

RESUMEN

We investigated the effect of blockade of muscarinic, cholinergic receptors by atropine on the Cd2(+)- and Ni2(+)-induced vasoconstriction of rings of endothelium-free, vascular smooth muscle from the ventral aorta of the dogfish shark, Squalus acanthias. Atropine reduced the Cd2(+)-induced vasoconstriction by approximately 50%, but did not alter the Ni2(+)-induced vasoconstriction, suggesting that the vasoactivity of these two metals may be, at least partially, via different pathways. In addition, this is the first demonstration that one component of Cd2+ vaso-toxicity may be via stimulation of muscarinic receptors.


Asunto(s)
Atropina/farmacología , Cadmio/toxicidad , Contracción Muscular/efectos de los fármacos , Músculo Liso Vascular/efectos de los fármacos , Níquel/toxicidad , Receptores Muscarínicos/efectos de los fármacos , Vasoconstricción/efectos de los fármacos , Animales , Aorta/efectos de los fármacos , Cazón , Receptores Muscarínicos/fisiología