Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
Crit Care Med ; 24(7): 1163-7, 1996 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-8674329

RESUMEN

OBJECTIVE: To compare point-of-care results obtained from an on-site hemocytometer with values provided by an institutional laboratory instrument. DESIGN: A prospective laboratory evaluation. SETTING: The central laboratory and cardiac surgical intensive care unit of a university-affiliated tertiary care center. PATIENTS: Normal range comparison was performed using blood specimens routinely obtained from 48 hospitalized patients for complete blood count analysis. The second evaluation was performed on blood specimens routinely obtained (in the intensive care unit) after cardiac surgery involving extracorporeal circulation in a series of 187 consecutive patients. MEASUREMENTS AND MAIN RESULTS: Hemoglobin concentration, platelet count, mean corpuscular volume, mean platelet volume, and red and white blood cell counts were measured with both on-site (MD 16, Coulter Electronics, Hialeah, FL) and laboratory (STKS, Coulter Electronics) instruments. Hematocrit and red cell distribution width were calculated using measured variables. Blood specimens were obtained from two distinct patients series. To evaluate measurement values within the normal range, a series of 48 routinely obtained blood specimens for complete blood count analysis in our institutional laboratory were utilized for concurrent analysis with the on-site hemocytometer. To evaluate measurement values out of the normal range, a second comparison involved measurements performed on blood specimens obtained in the cardiac surgical intensive care unit for complete blood count analysis. Linear regression demonstrated good correlations between on-site and laboratory hemoglobin concentration (r2 = .97), hematocrit (r2 = .95), platelet count (r2 = .97), mean corpuscular volume (r2 = .91), red cell distribution width (r2 = .80), and red (r2 = .95) and white (r2 = .96) blood cell count results. A marginal correlation was observed between mean platelet volume values (r2 = .47). Bias analysis (mean +/- 2 SD) demonstrated similar measurements between on-site and laboratory hemoglobin concentration, hematocrit, platelet count, red blood cell count, white blood cell count, mean platelet volume, mean corpuscular volume, and red cell distribution width. CONCLUSIONS: On-site hemoglobin concentration, hematocrit, white blood cell count, red blood cell count, red cell distribution width, and platelet count values compare well with those results obtained from the laboratory. The MD 16 hemocytometer (Coulter Electronics) provides on-site hematologic results that can provide an accurate and rapid quantitative assessment of platelets, and red and white blood cells. Rapid access to information obtained from this type of system may be clinically useful, especially in critically ill patients.


Asunto(s)
Pruebas Hematológicas/instrumentación , Pruebas Hematológicas/métodos , Recuento de Eritrocitos/instrumentación , Recuento de Eritrocitos/métodos , Índices de Eritrocitos , Circulación Extracorporea , Hematócrito/instrumentación , Hematócrito/métodos , Hemoglobinometría/instrumentación , Hemoglobinometría/métodos , Humanos , Unidades de Cuidados Intensivos , Laboratorios de Hospital , Recuento de Leucocitos/instrumentación , Recuento de Leucocitos/métodos , Recuento de Plaquetas/instrumentación , Recuento de Plaquetas/métodos , Estudios Prospectivos
2.
Anesth Analg ; 82(1): 13-21, 1996 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-8712388

RESUMEN

The purpose of this study was to prospectively evaluate whether heparin and protamine doses administered using a standardized protocol based on body weight and activated clotting time values are associated with either transfusion of hemostatic blood products (HBPs) or excessive postoperative bleeding. Analysis using 10 multiple logistic or linear regression models in 487 cardiac surgical patients included perioperative variables that may have an association with either transfusion of HBP and/or excessive postoperative chest tube drainage (CTD). Prolonged duration of cardiopulmonary bypass (CPB), lower pre-CPB heparin dose, lower core body temperature in the intensive care unit, combined procedures, older age, repeat procedures, a larger volume of salvaged red cells reinfused intraoperatively and abnormal laboratory coagulation results (prothrombin time, activated partial thromboplastin time, and platelet count) after CPB were associated with both transfusion of HBP and increased CTD. Female gender, lower total heparin dose, preoperative aspirin use and the number of HBPs administered intraoperatively were associated only with increased CTD, whereas a larger total protamine dose was associated only with perioperative transfusion of HBPs. Preoperative use of warfarin or heparin was not associated with excessive blood loss of perioperative transfusion of HBPs. In contrast to previous studies using bovine heparin, data from the present study do not support the use of reduced doses of porcine heparin during CPB.


Asunto(s)
Transfusión de Componentes Sanguíneos , Pérdida de Sangre Quirúrgica/prevención & control , Puente Cardiopulmonar/efectos adversos , Hemostasis Quirúrgica/métodos , Factores de Edad , Anciano , Animales , Tubos Torácicos , Drenaje , Femenino , Heparina/efectos adversos , Heparina/uso terapéutico , Humanos , Masculino , Persona de Mediana Edad , Análisis Multivariante , Estudios Prospectivos , Protaminas/efectos adversos , Protaminas/uso terapéutico , Factores de Riesgo , Porcinos , Factores de Tiempo , Tiempo de Coagulación de la Sangre Total