Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros










Base de datos
Intervalo de año de publicación
1.
Eur J Vasc Endovasc Surg ; 50(4): 480-6, 2015 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-26286386

RESUMEN

OBJECTIVE/BACKGROUND: To prospectively quantify radiation dose change in aortoiliac endovascular procedures in the hybrid operating room (OR) for patients and medical staff with a novel X-ray imaging technology (ClarityIQ technology), and to assess whether procedure or fluoroscopy time or dose of iodinated contrast was affected. METHODS: A prospective study including 138 patients was performed to compare radiation dose before and after installation of a novel X-ray imaging technology. Endovascular aneurysm repair (EVAR) was performed in 37 patients and an endovascular procedure for aortoiliac occlusive disease (AIOD) in 101. Patient radiation dose in air kerma (AK) and dose area product (DAP), patient demographics, and procedural data were recorded. Staff radiation dose was measured with real time personal dosimetry measurements. In both the EVAR and AIOD groups the reference system, ALX (AlluraXper FD20; Philips Healthcare, Best, the Netherlands), was compared with the upgraded X-ray system, CIQ (AlluraClarity FD20; Philips Healthcare). Procedure time, fluoroscopy time, and iodinated contrast dose were recorded. RESULTS: Patient radiation dose reduction in the EVAR group, in median AK, was 56% (ALX = 1,262.5 mGy; CIQ = 556.0 mGy [p < .01]); and in median DAP it was 57% (ALX = 224.4 Gycm(2) and CIQ = 95.8 Gycm(2) [p < .01]). Patient radiation dose reduction in the AIOD group, in median AK, was 76% (ALX = 1,011.0 mGy; CIQ = 248.0 mGy [p < .01]); and in median DAP it was 73% (ALX = 138.1 Gycm(2); CIQ = 38.0 Gycm(2) [p < .01]). Staff dose reduction in the EVAR group was 16% (ALX = 70.1 µSv; CIQ = 59.2 µSv [p = .43]) and in the AIOD group it was 69% (ALX = 96.2 µSv; CIQ = 30.1 µSv [p < .01]). There was no statistically significant difference between patient demographics, procedure time, fluoroscopy time, and iodinated contrast medium use in the two treatment groups before and after installation. CONCLUSION: A novel X-ray imaging technology in the hybrid OR suite resulted in a significant reduction of patient and staff radiation dose without affecting procedure length, fluoroscopy time, or use of contrast.


Asunto(s)
Aneurisma de la Aorta/terapia , Aortografía/instrumentación , Arteriopatías Oclusivas/terapia , Procedimientos Endovasculares/instrumentación , Quirófanos , Dosis de Radiación , Radiografía Intervencional/instrumentación , Anciano , Aneurisma de la Aorta/diagnóstico por imagen , Aortografía/efectos adversos , Arteriopatías Oclusivas/diagnóstico por imagen , Medios de Contraste/administración & dosificación , Procedimientos Endovasculares/efectos adversos , Diseño de Equipo , Femenino , Humanos , Masculino , Persona de Mediana Edad , Exposición Profesional , Estudios Prospectivos , Traumatismos por Radiación/etiología , Traumatismos por Radiación/prevención & control , Monitoreo de Radiación/instrumentación , Protección Radiológica/instrumentación , Radiografía Intervencional/efectos adversos , Factores de Riesgo , Factores de Tiempo , Resultado del Tratamiento
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...