Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 17 de 17
Filtrar
1.
Cancer Radiother ; 27(4): 303-311, 2023 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-37149464

RESUMO

PURPOSE: This study presents the methodology and results of the acceptance and periodical quality controls on the MRIdian®. MATERIALS AND METHODS: The impact of the magnetic field on other machines was investigated by controlling nearby linacs dose profiles. The image quality of the 0.345T MR scanner was evaluated, also assessing the integrated linear accelerator influence. The photon beams lateral and depth dose profiles were measured in motorized water tanks, along dose rate and output factors, and compared to Monte Carlo (MC) calculations. The isocenter position, gantry angles and multi-leaf collimator (MLC) position were controlled using film dosimetry. Gating latency and dosimetric accuracy were controlled with a dynamic phantom. RESULTS: The magnetic field had no significant impact on other nearby linacs. Image quality was within tolerances and did not vary over time. Dose profiles measured showed good agreement with MC data, with maximum differences of 1.3% in-field. Output factors were within 0.8% of calculated values. Imaging and radiative isocenter matched within 0.9±0.4mm over all monthly controls. Gantry rotation was precise within -0.1±0.2°, with an isocenter variation of 1.4±0.3mm diameter. The average MLC position was within 0.4±0.1mm of theoretical value. Finally, the gating latency was 0.14±0.07sec and the gated dose within 0.3% of base value. CONCLUSION: All results are within the tolerances fixed by ViewRay® and show low variations over 2 years, comforting the use of small margins and gating for high-dose adaptive treatments.


Assuntos
Fótons , Radiometria , Humanos , Seguimentos , Radiometria/métodos , Rotação , Imagens de Fantasmas , Aceleradores de Partículas , Planejamento da Radioterapia Assistida por Computador
2.
Cancer Radiother ; 27(5): 355-361, 2023 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-37085341

RESUMO

PURPOSE: Adaptive radiotherapy with the Ethos® therapy Varian system has been recently implemented at the Montpellier Cancer Institute, France. This article details the commissioning performed before the implementation of this new treatment planning system (TPS). MATERIAL AND METHODS: To validate the golden beam data of the machine (Halcyon linear accelerator), percentage depth doses (PDD) and profiles were measured for several field sizes and at different depths with a microdiamond chamber. The final doses calculated for different plan types with the Ethos Acuros XB algorithm and the Halcyon Eclipse Analytic Anisotropic Algorithm were compared using the gamma index method. Lastly, for the patient quality assurance (QA) process, the patient treatment plan results obtained with the Mobius3D QA platform (Varian) were compared with the portal dosimetry results obtained with Epiqa (Epidos). RESULTS: Minor differences were observed for the PDD and profile curves (mean difference of 0.2% and 2%, respectively). The χ index pass rate was above 98% for all measures using the 1%/1mm and 2%/2mm criteria for PDD and profile evaluations. The Ethos AXB algorithm was validated for every configuration (fixed fields, standard IMRT and VMAT fields, and clinical plans) with 2D/3D gamma index values>99%. Seventy-three 3-arcs-VMAT QA plans and 27 9-fields-IMRT QA plans were evaluated. Both showed excellent agreement with the TPS calculations (mean gamma pass rate higher than 99%). No difference was observed between IMRT and VMAT. CONCLUSION: The beam delivery, the Ethos AXB algorithm, and the patient QA were comprehensively validated using independent tools.


Assuntos
Planejamento da Radioterapia Assistida por Computador , Radioterapia de Intensidade Modulada , Humanos , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Radiometria , Algoritmos , Aceleradores de Partículas
3.
Cancer Radiother ; 23(6-7): 636-650, 2019 Oct.
Artigo em Francês | MEDLINE | ID: mdl-31444078

RESUMO

Liver stereotactic body radiotherapy is a developing technique for the treatment of primary tumours and metastases. Its implementation is complex because of the particularities of the treated organ and the comorbidities of the patients. However, this technique is a treatment opportunity for patients otherwise in therapeutic impasse. The scientific evidence of liver stereotactic body radiotherapy has been considered by the French health authority as insufficient for its widespread use outside specialized and experienced centers, despite a growing and important number of retrospective and prospective studies, but few comparative data. This article focuses on the specific features of stereotactic body radiotherapy for liver treatments and the results of published studies of liver stereotactic body radiotherapy performed with classic linear accelerators and dedicated radiosurgery units.


Assuntos
Carcinoma Hepatocelular/radioterapia , Neoplasias Hepáticas/radioterapia , Radiocirurgia/instrumentação , Carcinoma Hepatocelular/diagnóstico por imagem , Humanos , Fígado/efeitos da radiação , Cirrose Hepática/complicações , Neoplasias Hepáticas/diagnóstico por imagem , Neoplasias Hepáticas/secundário , Movimentos dos Órgãos , Aceleradores de Partículas , Guias de Prática Clínica como Assunto , Tolerância a Radiação , Radiocirurgia/métodos , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia Guiada por Imagem/métodos , Respiração , Resultado do Tratamento
4.
Cancer Radiother ; 22(1): 25-30, 2018 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-29221676

RESUMO

PURPOSE: Radiation oncologists are responsible for deciding which day-to-day variations are acceptable or not in the treatment setup. However, properly qualified and trained radiation therapists might be capable to perform image registration. We evaluated in our centre the capability and accuracy of radiation therapists to validate positioning images in a prospective study. METHODS AND PATIENTS: A total of 84 patients treated for prostate, head and neck, lung or breast cancer was prospectively and randomly included from July 2011 to July 2013 in radiotherapy unit of our institution. For each patient, three positioning images were randomly analysed. Two radiation oncologists analysed all positioning images and shifts decided by the radiation therapists in an independent and blinded way. The radiation oncologists had to decide whether to validate or not this shift and give a corresponding additional shift, if any. A theoretical disagreement rate less than 5% between radiation therapists and radiation oncologists was planned. RESULTS: A total of 240 images were analysed (head and neck: 15.0%; prostate: 14.2%; breast: 55.0%; lung: 15.8%). The global disagreement between radiation oncologists and radiation therapists for all the images analysed was 2.5% 95% confidence interval (95% CI) [1.0-5.0], corresponding to six images out of 240. A 100% agreement was reached for prostate and lung images, a 97.2% agreement for head and neck images and a 96.2% agreement for breast images. CONCLUSIONS: The radiation therapist validation for repositioning images seemed accurate for image-guided radiotherapy in our institution. Periodic evaluation and in-house training are warranted when routine delegation of image registration to radiation therapists is considered.


Assuntos
Pessoal Técnico de Saúde , Neoplasias/radioterapia , Variações Dependentes do Observador , Posicionamento do Paciente , Radio-Oncologistas , Radioterapia Guiada por Imagem , Feminino , França , Humanos , Masculino , Estudos Prospectivos
5.
Cancer Radiother ; 20 Suppl: S196-9, 2016 Sep.
Artigo em Francês | MEDLINE | ID: mdl-27521030

RESUMO

Surgery (radical cystectomy) is the standard treatment of muscle-invasive bladder cancer. Radiochemotherapy has risen as an alternative treatment option to surgery as part as organ-sparing combined modality treatment or for patients unfit for surgery. Radiochemotherapy achieves 5-year bladder intact survival of 40 to 65% and 5-year overall survival of 40 to 50% with excellent quality of life. This article introduces the French recommendations for radiotherapy of bladder cancer: indications, exams, technique, dosimetry, delivery and image guidance.


Assuntos
Carcinoma de Células de Transição/radioterapia , Neoplasias da Bexiga Urinária/radioterapia , Carcinoma de Células de Transição/cirurgia , Terapia Combinada , Cistectomia , Fracionamento da Dose de Radiação , Humanos , Irradiação Linfática , Metástase Linfática , Órgãos em Risco , Lesões por Radiação/prevenção & controle , Radioterapia/efeitos adversos , Radioterapia/métodos , Radioterapia/normas , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Radioterapia Adjuvante/métodos , Radioterapia Guiada por Imagem/métodos , Neoplasias da Bexiga Urinária/cirurgia
6.
Radiat Oncol ; 11: 2, 2016 Jan 14.
Artigo em Inglês | MEDLINE | ID: mdl-26762179

RESUMO

PURPOSE: To verify the accuracy of volumetric arc therapy (VMAT) using the RapidArc™ device when switching patients from one single linear accelerator (linac) to a paired energy and mechanics "twin" linac without reoptimization of the original treatment plan. PATIENTS AND METHODS: Four centers using 8 linacs were involved in this study. Seventy-four patients previously treated with the 6MV photon RapidArc™ technique were selected for analysis, using 242 measurements. In each institution, all patients were planned on linac A, and their plans were verified both on linac A and on the twin linac B. Verifications were done using the amorphous silicium electronic portal imager (EPID) of the linacs and were analyzed with the EpiQa software (Epidos, Bratislavia, Slovakia). The gamma index formalism was used for validation with a double threshold of 3 % and 3 mm with a measurement resolution of 0.39 mm/pixel, and a smoothed resolution of approximately 2.5 mm. RESULTS: The number of points passing the gamma criteria between the measured and computed doses was 94.79 ± 2.57 % for linac A and 94.61 ± 2.46 % for linac B. Concerning the smoothed measurement analysis, 98.67 ± 1.26 % and 98.59 ± 1.20 % points passing the threshold were obtained for linacs A and B, respectively. The difference between the 2 dose matrices acquired on the EPID was very small, with 99.92 ± 0.06 % of the points passing the criteria. CONCLUSION: For linacs sharing the same mechanical and energy parameters, this study tends to indicate that patients may be safely switched from treatment with one linac to treatment with its twin linac using the same VMAT plan.


Assuntos
Neoplasias de Cabeça e Pescoço/radioterapia , Neoplasias/radioterapia , Neoplasias Pélvicas/radioterapia , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia Guiada por Imagem/métodos , Radioterapia de Intensidade Modulada/métodos , Algoritmos , Calibragem , Eletrônica , Feminino , França , Humanos , Masculino , Distribuição Normal , Aceleradores de Partículas , Garantia da Qualidade dos Cuidados de Saúde , Interpretação de Imagem Radiográfica Assistida por Computador/métodos , Software
7.
Cancer Radiother ; 19(6-7): 532-7, 2015 Oct.
Artigo em Francês | MEDLINE | ID: mdl-26344440

RESUMO

The ultimate goal in radiation oncology is to offer a personalized treatment to all patients indicated for radiotherapy. Radiomics is a tool that reinforces a deep analysis of tumors at the molecular aspect taking into account intrinsic susceptibility in a long-term follow-up. Radiomics allow qualitative and quantitative performance analyses with high throughput extraction of numeric radiologic data to obtain predictive or prognostic information from patients treated for cancer. A second approach is to define biological or constitutional that could change the practice. This technique included normal tissue individual susceptibility but also potential response of tumors under ionizing radiation treatment. These "omics" are biological and technical techniques leading to simultaneous novel identification and exploration a set of genes, lipids, proteins.


Assuntos
Radioterapia/métodos , Genômica , Humanos , Neoplasias/radioterapia , Proteômica , Radiação Ionizante
8.
Cancer Radiother ; 14 Suppl 1: S161-73, 2010 Nov.
Artigo em Francês | MEDLINE | ID: mdl-21129660

RESUMO

Radiation therapy is now widely accepted as an efficacious treatment of localized prostate cancer. The technical developments of recent years have enabled the evolution of a three-dimensional conformal radiotherapy, offering a better adaptation of the dose distribution, and leading therefore to preserve organs at risk. In addition, the required dose delivered to the target volume permit physician to increase the total dose if necessary. This requires a thorough knowledge of the radio-anatomy of the prostate, the natural history of the disease but also the ballistics and dosimetry. The objectives of this work were to detail epidemiology and radio-anatomy of the prostate cancer. In addition, conformal radiation modalities are illustrated by a case report.


Assuntos
Neoplasias da Próstata/radioterapia , Radioterapia Conformacional/métodos , Braquiterapia/métodos , Humanos , Masculino , Próstata/anatomia & histologia , Próstata/efeitos da radiação , Neoplasias da Próstata/epidemiologia , Neoplasias da Próstata/patologia , Dosagem Radioterapêutica/normas , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/métodos
9.
Cancer Radiother ; 14 Suppl 1: S174-81, 2010 Nov.
Artigo em Francês | MEDLINE | ID: mdl-21129661

RESUMO

More than one third of patients with localized prostate cancer at diagnosis who receive radical prostatectomy have histologically extraprostatic disease. Three randomized trials have demonstrated a significant benefit of postoperative radiotherapy for these patients in terms of biochemical progression-free survival, overall survival or metastasis-free survival in function of each study, at the cost of moderate acute and late toxicity. The technical developments of recent years have enabled the evolution of a three-dimensional conformal radiotherapy, with better adaptation of the dose distribution to the shape of target volumes to preserve organs at risk while delivering the required dose volume target. This requires a thorough knowledge of the prostate cancer radioanatomy, ballistics and dosimetry. Purpose of this work was to specify epidemiological and radioanatomy characteristics for this tumor type and conformal radiation modalities illustrated by a case report.


Assuntos
Neoplasias da Próstata/radioterapia , Radioterapia Conformacional/métodos , Terapia Combinada/métodos , Intervalo Livre de Doença , Humanos , Incidência , Masculino , Período Pós-Operatório , Próstata/anatomia & histologia , Próstata/efeitos da radiação , Prostatectomia , Neoplasias da Próstata/epidemiologia , Neoplasias da Próstata/patologia , Neoplasias da Próstata/cirurgia , Dosagem Radioterapêutica , Radioterapia Conformacional/efeitos adversos
10.
Cancer Radiother ; 14 Suppl 1: S52-60, 2010 Nov.
Artigo em Francês | MEDLINE | ID: mdl-21129670

RESUMO

Cancers of the paranasal sinuses are rare tumors, with treatment based on a multidisciplinary approach. Surgery and radiation therapy, possibly associated with chemotherapy are used to obtain 5 years specific survival rate of 60-70 %. Advances in radiotherapy, including the use of imaging for 3D conformal approach require precise knowledge of the radioanatomy for this type of tumor to determine the different volumes of interest. Purpose of this study was to specify radioanatomy and conformal radiation modalities for cancers of the sinuses, and is illustrated by a case report.


Assuntos
Neoplasias dos Seios Paranasais/radioterapia , Seios Paranasais/anatomia & histologia , Antineoplásicos/uso terapêutico , Terapia Combinada , Humanos , Neoplasias dos Seios Paranasais/tratamento farmacológico , Neoplasias dos Seios Paranasais/epidemiologia , Neoplasias dos Seios Paranasais/cirurgia , Seios Paranasais/diagnóstico por imagem , Radiografia , Radioterapia Conformacional/métodos
11.
Cancer Radiother ; 14(6-7): 550-3, 2010 Oct.
Artigo em Francês | MEDLINE | ID: mdl-20724189

RESUMO

Intensity modulated radiation therapy (IMRT) offers optimal dosimetric and clinical results in terms of acute toxicity, allows augmenting the dose to the target volumes and therefore, appears promising for local control and disease-free survival. However, several pitfalls to this treatment are to be considered, namely a long treatment time and a high number of monitor unit (MU) required. The dosimetric results of the volumetric modulated arctherapy gives at least similar target coverage and preservation of organs at risk, while significantly reducing the number of required MUs and the overall treatment time. This has a potential impact on the treatment quality and the potential risk of secondary cancers. Volumetric modulated arctherapy allows implementation of stereotactic radiation therapy and complex treatments previously considered not feasible with IMRT. The future will involve this technology of high precision to determine the dose and to the target in real time using the image-guided radiotherapy. Tools combining these two methods are in development.


Assuntos
Neoplasias/radioterapia , Radioterapia (Especialidade) , Radioterapia de Intensidade Modulada/métodos , Antineoplásicos Hormonais/uso terapêutico , Neoplasias do Ânus/radioterapia , Ensaios Clínicos como Assunto , Terapia Combinada , Sistemas Computacionais , Tomografia Computadorizada de Feixe Cônico , Relação Dose-Resposta à Radiação , Feminino , Neoplasias dos Genitais Femininos/radioterapia , Humanos , Intestinos/efeitos da radiação , Masculino , Neoplasias/diagnóstico por imagem , Neoplasias/patologia , Neoplasias/cirurgia , Neoplasias Induzidas por Radiação/etiologia , Neoplasias Induzidas por Radiação/prevenção & controle , Segunda Neoplasia Primária/etiologia , Segunda Neoplasia Primária/prevenção & controle , Neoplasias da Próstata/diagnóstico por imagem , Neoplasias da Próstata/tratamento farmacológico , Neoplasias da Próstata/radioterapia , Lesões por Radiação/etiologia , Lesões por Radiação/prevenção & controle , Radiocirurgia , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/efeitos adversos , Radioterapia de Intensidade Modulada/instrumentação , Radioterapia de Intensidade Modulada/normas , Risco , Bexiga Urinária/efeitos da radiação
12.
Bull Cancer ; 97(7): 769-78, 2010 Jul.
Artigo em Francês | MEDLINE | ID: mdl-20610365

RESUMO

Intensity-modulated radiotherapy (IMRT) has highly impacted on the dose delivery thanks to inverse-planning dosimetry. Conformal isodoses to target volumes and critical organ protection have led to treatment possibilities which were unrealizable with conventional 3D technique. Nevertheless, time delivery using IMRT was in some cases longer than 3D radiotherapy. In order to compensate for this limitation, we have developed since 2007 a volumetric modulated arctherapy (RapidArc) in partnership with Varian. The technique, the results of the dosimetry plans, and quality control of the 142 first patients treated since November 2008 are presented in this review.


Assuntos
Neoplasias do Ânus/radioterapia , Neoplasias dos Genitais Femininos/radioterapia , Neoplasias da Próstata/radioterapia , Radioterapia de Intensidade Modulada/métodos , Sarcoma/radioterapia , Algoritmos , Institutos de Câncer , Feminino , França , Humanos , Masculino , Lesões por Radiação/prevenção & controle , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/instrumentação
13.
Cancer Radiother ; 13(5): 409-15, 2009 Sep.
Artigo em Francês | MEDLINE | ID: mdl-19246230

RESUMO

Radiation therapy is now widely accepted as an efficacious treatment of localized prostate cancer. Recent advances, namely with the development of conformal radiotherapy, allowed to increase the total dose in the target volumes with greater local control. A forward step achieved with intensity modulated radiotherapy (IMRT) in terms of therapeutic ratio between target volumes and critical organs. IMRT offers an inverse planning dosimetry and a modulation of the fields during irradiation. This article presents recent technical and clinical advances in IMRT focused on prostate cancer.


Assuntos
Neoplasias da Próstata/radioterapia , Radioterapia de Intensidade Modulada/normas , Humanos , Masculino , Movimento , Neoplasias da Próstata/diagnóstico por imagem , Neoplasias da Próstata/patologia , Radiografia , Dosagem Radioterapêutica , Radioterapia de Intensidade Modulada/métodos , Radioterapia de Intensidade Modulada/tendências , Reto/efeitos da radiação , Glândulas Seminais/efeitos da radiação , Bexiga Urinária/efeitos da radiação
14.
Cancer Radiother ; 8(2): 59-69, 2004 Apr.
Artigo em Inglês, Francês | MEDLINE | ID: mdl-15063872

RESUMO

PURPOSE: - To report our experience on treatment planning and acute toxicity in 16 patients suffering from clinically localized prostate cancer treated with high-dose intensity-modulated radiation therapy (IMRT). PATIENTS AND METHODS: - Between March 2001 and October 2002, 16 patients with clinically localized prostate cancer were treated with IMRT. Treatment planning included an inverse-planning approach, and the desired beam intensity profiles were delivered by dynamic multileaf collimation. All patients received the entire treatment course with IMRT to a prescribed dose of 78 Gy. All IMRT treatment plans were compared with a theoretical conventional three-dimensional conformal radiation therapy (3D-CRT). Acute lower gastro-intestinal (GI) and genito-urinary (GU) toxicity was evaluated in all patients and graded according to the Common Toxicity Criteria for Adverse Events version 3.0 (CTCAE v. 3.0). A relationship between dose volume and clinical toxicity was evaluated. RESULTS: - Ninety-five percent of the PTV2 received more than 76 Gy using IMRT or 3D-CRT with no difference between both methods. The dose-volume histogram mean obtained for the PTV2 was not different between IMRT and 3D-CRT. IMRT improved homogeneity of the delivered dose to the PTV2 as compared with 3D-CRT (7.5 vs 9%, respectively). Ninety-five percent of the PTV1 received 5 Gy more using IMRT with protection of the bladder and the rectum walls. The benefit was considered below 75 and 70 Gy for the wall of the bladder and the rectum, respectively. Grade 2 GI and GU toxicity was observed in four (25%) and five (31%) patients, respectively. No grade 3 toxicity was observed. There was a trend towards a relationship between the mean rectal dose and acute rectal toxicity but without statistical significant difference (P =0.09). CONCLUSION: - Dose escalation with IMRT is feasible with no grade 3 or higher acute GI or GU toxicity. Examination of a larger cohort and longer-term follow-up are warranted in the future.


Assuntos
Adenocarcinoma/radioterapia , Neoplasias da Próstata/radioterapia , Radioterapia Conformacional/métodos , Adenocarcinoma/patologia , Idoso , Sistema Digestório/efeitos da radiação , Humanos , Imageamento Tridimensional , Masculino , Pessoa de Meia-Idade , Estadiamento de Neoplasias , Projetos Piloto , Próstata/patologia , Neoplasias da Próstata/patologia , Doses de Radiação , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Radioterapia Conformacional/efeitos adversos , Reto/efeitos da radiação , Sistema Urogenital/efeitos da radiação
15.
Cancer Radiother ; 8 Suppl 1: S121-7, 2004 Nov.
Artigo em Francês | MEDLINE | ID: mdl-15679257

RESUMO

INTRODUCTION: Between May 2002 and May 2004, eight French comprehensive cancer centres did a prospective nonrandomized study including 200 patients, 100 with cancer of the prostate and 100 with head and neck cancers. Half of each patient group was treated by IMRT and the others by RTC 3D. This clinical study was associated with an economic study and a physics study. We report here the first results. PATIENTS AND METHODS: For the clinical study, the analysis of the data of the first 88 patients irradiated for a prostatic cancer shows that 39 received RTC and 49 IMRT with a mean dose of 78 Gy at the ICRU point at 2 Gy per fraction. For H&N tumours, the preliminary analysis was done on the 87 first patients with a mean follow-up of 11.5 months (2 to 25 months) and a median of 8.4 months for the IMRT groups and 13.2 months for the RTC group. The economic study was done on the first 157 patients included during the first 18 months: 71 treated by RTC (35 for H&N and 36 for prostate) and 86 treated by IMRT (38 for H&N and 48 for prostate). The assessment of the direct costs was realized by a micro-costing technique. The physical study compared dose distributions for both techniques and has created quality control recommendations. RESULTS: Clinical studies of the acute reactions do not show any difference between groups, but we want to point out the short follow-up and the relatively high dose delivered to cancers of the prostate. The physics study demonstrates that IMRT is technically feasible in good clinical conditions with high quality assurance, a good reproducibility and precision. Dosimetric data show that IMRT could certainly spare organs at risk more than RTC for H&N tumours. The direct costs of "routine" treatments for H&N tumours were 4922 euros for IMRT versus 1899 euros for RTC and for the prostatic cancers 4911 euros for IMRT versus 2357 for RTC.


Assuntos
Neoplasias de Cabeça e Pescoço/radioterapia , Neoplasias da Próstata/radioterapia , Planejamento da Radioterapia Assistida por Computador , Radioterapia Conformacional/métodos , Adolescente , Adulto , Idoso , Idoso de 80 Anos ou mais , Quimioterapia Adjuvante , Terapia Combinada , Análise Custo-Benefício , Seguimentos , Neoplasias de Cabeça e Pescoço/tratamento farmacológico , Neoplasias de Cabeça e Pescoço/mortalidade , Neoplasias de Cabeça e Pescoço/cirurgia , Humanos , Masculino , Pessoa de Meia-Idade , Neoplasias Otorrinolaringológicas/mortalidade , Neoplasias Otorrinolaringológicas/radioterapia , Estudos Prospectivos , Neoplasias da Próstata/mortalidade , Dosagem Radioterapêutica , Radioterapia Conformacional/economia , Fatores de Tempo
16.
Br J Cancer ; 89(10): 1987-94, 2003 Nov 17.
Artigo em Inglês | MEDLINE | ID: mdl-14612914

RESUMO

The aim of this study was to treat carcinoembryonic antigen (CEA)-expressing pancreatic carcinoma cells with tumour necrosis factor alpha (TNFalpha) and simultaneous radiation therapy (RT), using a bispecific antibody (BAb) anti-TNFalpha/anti-CEA. TNFalpha used alone produced a dose-dependent inhibition of the clonogenic capacity of the cultured cells. Flow cytometry analysis of cell cycle progression confirmed the accumulation of cells in G(1) phase after exposure to TNFalpha. When TNFalpha was added 12 h before RT, the surviving fraction at 2 Gy was 60% lower than that obtained with irradiation alone (0.29 vs 0.73, respectively, P<0.00001). In combination treatment, cell cycle analysis demonstrated that TNFalpha reduced the number of cells in radiation-induced G(2) arrest, blocked irreversibly the cells in G(1) phase, and showed an additive decrease of the number of cells in S phase. In mice, RT as a single agent slowed tumour progression as compared with the control group (P<0.00001). BAb+TNFalpha+RT combination enhanced the delay for the tumour to reach 1500 mm(3) as compared with RT alone or with RT+TNFalpha (P=0.0011). Median delays were 90, 93, and 142 days for RT alone, RT+TNFalpha, and RT+BAb+TNFalpha groups, respectively. These results suggest that TNFalpha in combination with BAb and RT may be beneficial for the treatment of pancreatic cancer in locally advanced or adjuvant settings.


Assuntos
Adenocarcinoma/patologia , Anticorpos Biespecíficos/imunologia , Antineoplásicos/farmacologia , Neoplasias Pancreáticas/patologia , Fator de Necrose Tumoral alfa/farmacologia , Antineoplásicos/imunologia , Antígeno Carcinoembrionário/análise , Antígeno Carcinoembrionário/biossíntese , Humanos , Tolerância a Radiação , Radiação Ionizante , Fase S , Células Tumorais Cultivadas , Fator de Necrose Tumoral alfa/imunologia
17.
Med Phys ; 23(8): 1421-4, 1996 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-8873040

RESUMO

A new system of radiation dose mapping based on laser heating of thermoluminescent dosimetric plates (TLDP) has been developed. Application of this technique to intraoperative radiotherapy (IORT) has been investigated. Preliminary results show that TLDP are a possible alternative to radiological films and ionization chambers since they provide some advantages such as 2D dose measurements with digital storage, large dynamic dose range, and easy processing.


Assuntos
Neoplasias/diagnóstico por imagem , Neoplasias/cirurgia , Radioterapia/instrumentação , Radioterapia/métodos , Terapia Combinada , Elétrons , Temperatura Alta , Humanos , Lasers , Medições Luminescentes , Imagens de Fantasmas , Radiografia , Dosagem Radioterapêutica
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...