Ultra-high molecular weight polyethylene wear particles have been implicated as the major cause of osteolysis, implant loosening and late aseptic failure in total hip arthroplasties in vivo. This study initially screened 22 carbon-carbon composite materials as alternatives for UHMWPE in joint bearings. New bearing materials should satisfy certain criteria--they should have good wear properties that at least match UHMWPE, and produce wear particles with low levels of cytotoxic and osteolytic activity. Initial screening was based on wear resistance determined in short-term tribological pin-on-plate tests. Three materials (HMU-PP(s), HMU-RC-P(s), and SMS-RC-P(s)) which had superior wear resistance were selected for long-term testing. All materials had very low wear factors and SMS-RC-P(s), which had a wear factor of 0.08 +/- 0.56 x 10(-7) mm3/Nm, was selected for the subsequent biological testing and particle size analysis. SMS-RC-P(s) showed good biocompatibility in bulk material form and also the wear particles had low cytotoxicity for L929 fibroblasts in culture compared to metal wear particles. Wear debris size analysis by transmission electron microscopy showed that the particles were very small, with the vast majority being under 100 nm in size, similar to metal wear particles. The potential osteolytic effect of SMS-RC-P(s) wear particles was investigated by culturing particles with human peripheral blood mononuclear cells and measuring TNFalpha production. SMS-RC-P(s) did not significantly stimulate TNFalpha production at a particle volume to cell number ratio of 80:1, indicating that the debris had a low osteolytic potential. The results of this study suggest that carbon-carbon composites, particularly those composed of PAN-based fibers may be important biomaterials in the development of next generation bearing surfaces for use in total joint replacements that have very low wear rates and reduced osteolytic and cytotoxic potential.


Assuntos
Materiais Biocompatíveis , Carbono , Prótese de Quadril , Animais , Materiais Biocompatíveis/toxicidade , Linhagem Celular , Sobrevivência Celular/efeitos dos fármacos , Análise de Falha de Equipamento , Humanos , Técnicas In Vitro , Leucócitos Mononucleares/imunologia , Teste de Materiais , Camundongos , Microscopia Eletrônica , Tamanho da Partícula , Falha de Prótese , Fator de Necrose Tumoral alfa/biossíntese
2.
J Biomed Mater Res B Appl Biomater ; 67(2): 758-64, 2003 Nov 15.
Artigo em Inglês | MEDLINE | ID: mdl-14598403

RESUMO

UHMWPE wear particles have been implicated in osteolysis, implant loosening, and long-term failure of total hip arthroplasties in vivo. This study examined four carbon-based composite materials as alternatives for UHMWPE in joint bearings. These materials were HMU-CVD, SMS-CVD, P25-CVD, and CFR-PEEK. New bearing materials should satisfy certain criteria: they should have good wear properties that at least match UHMWPE, and produce wear particles with low levels of biological activity. Of the four materials tested in multidirectional pin-on-plate tribological tests, SMS-CVD, P25-CVD, and CFR-PEEK showed lower volumetric wear factors than UHMWPE. P25-CVD had the lowest wear factor of 0.54 +/- 0.34 x 10(-7) mm(3)/Nm. Analysis of P25-CVD wear particles by transmission electron microscopy showed that the debris was very small, with the vast majority of particles being under 100 nm in size, which was similar in size to metal wear particles. The P25-CVD particles were isolated and cultured with L929 fibroblasts and U937 monocytic cells to assess their effect on cell viability. P25-CVD particles were significantly less cytotoxic (p < 0.01, ANOVA) to both cell types than CoCr metal wear particles. This work suggests that carbon-carbon composite materials may have potential for use in total hip replacement bearings. Of the materials tested P25-CVD had the lowest wear factor, and produced very small wear debris that had minimal cytotoxic effect on L929 and U937 cells in vitro. Therefore carbon-carbon composites, such as P25-CVD, may be important in the development of next-generation implants with lower wear rates and reduced cytotoxic potential.


Assuntos
Materiais Biocompatíveis/normas , Sobrevivência Celular , Resinas Compostas/química , Prótese de Quadril/normas , Animais , Materiais Biocompatíveis/química , Carbono , Técnicas de Cultura de Células , Linhagem Celular , Cromo/farmacologia , Cobalto/farmacologia , Resinas Compostas/normas , Fibroblastos , Humanos , Teste de Materiais , Camundongos , Tamanho da Partícula , Propriedades de Superfície , Células U937