1.
Anal Sci ; 29(7): 753-5, 2013.
Artigo em Inglês | MEDLINE | ID: mdl-23842420

RESUMO

A definitive method is described for the indirect assay of several tens of milligrams of urea by coulometric titration. Urea was decomposed in concentrated sulfuric acid using a Kjeldahl flask. Subsequently, the formed ammonium ion was titrated with electrogenerated hypobromite ion in a sodium bromide-sodium tetraborate medium of pH 8.6, with amperometric end-point detection. Parameters affecting the pretreatment procedure were evaluated. The optimized conditions included the heating of 2 g of urea at around 300°C for 2 h with 10 cm(3) of sulfuric acid. Under the proposed conditions, the assay value with expanded uncertainty (k = 2), 99.870 ± 0.026%, agreed well with the certified value of NIST SRM 912a urea, 99.9 ± 0.1%.


Assuntos
Bromatos/química , Técnicas Eletroquímicas , Ureia/análise
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...