1.
J Arthroplasty ; 37(6S): S301-S305, 2022 06.
Artigo em Inglês | MEDLINE | ID: mdl-35219817

RESUMO

BACKGROUND: Prosthetic joint infections have become the leading cause of joint replacement failure. The primary sources of contamination are skin flora and bacteria from airborne particles. Portable ultraviolet air disinfection units are used in the Operating Room (OR) to prevent contamination from airborne particles; however, their effectiveness is not proven. The purpose of this study was to compare the rate of contamination of sites with and without Ultraviolet (UV) air disinfection units during active surgeries. METHODS: Sedimentation rates of viable particles were measured during 40 primary TKA procedures. Half of the procedures were performed with ultraviolet air disinfection units. Air-borne particles were collected on nitrocellulose membranes at 5 locations within the OR. After incubation, all microbial colonies were counted and the sedimentation rates were reported in CFUs/m2/hr. 10 additional trials were performed in an empty OR with no staff present. RESULTS: The average contamination rate of all sites was 22 ± 1.1 CFUs/m2/hr in the empty OR vs. 21.3 ± 4.6 CFUs/m2/hr with UV units and 20.3 ± 4.9 CFUs/m2/hr without (P = .03, P = .03, P = .964). Viable contaminates were found in the sterile field in 25% of UV cases vs 45% non-UV. These differences were not statistically significant. There were differences found however, according to the number of staff in the room (6 vs 7 staff: P = .036, 6 vs 8 staff: P = .004). CONCLUSION: There was no statistical difference in contamination rate with the usage or non-usage of UV units. These 40 cases shows that the largest variables affecting the contamination rate were the number of staff present and size of the OR.


Assuntos
Artroplastia do Joelho , Desinfecção , Microbiologia do Ar , Bactérias , Desinfecção/métodos , Humanos , Salas Cirúrgicas , Raios Ultravioleta
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...