Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Appl Radiat Isot ; 178: 109974, 2021 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-34653881

RESUMO

Knowledge of a radiation detector's numerical characteristics allows its energy efficiency to be calibrated theoretically for any measurement geometry. Here, energy efficiency calibration is discussed for oriented LaBr3(Ce) detectors used for in situ radiation monitoring. Vertical and horizontal detector orientations relative to the ground surface are compared using efficiency calibrations based on the detector's numerical characteristics and are equally effective. Quantitative assessment of the different measurement geometries was performed using a new analytical approach: integrated absolute full energy peak efficiency.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...