1.
J Clin Microbiol ; 54(11): 2763-2766, 2016 11.
Artigo em Inglês | MEDLINE | ID: mdl-27582513

RESUMO

We compared two rapid, point-of care nucleic acid amplification tests for detection of influenza A and B viruses (Alere i [Alere] and cobas Liat [Roche Diagnostics]) with the influenza A and B virus test components of the FilmArray respiratory panel (BioFire Diagnostics) using 129 respiratory specimens collected in universal viral transport medium (80 influenza A virus and 16 influenza B virus positive) from both adult and pediatric patients. The sensitivities of the Alere test were 71.3% for influenza A virus and 93.3% for influenza B virus, with specificities of 100% for both viruses. The sensitivities and specificities of the Liat test were 100% for both influenza A and B viruses. The poor sensitivity of the Alere test for detection of influenza A virus was likely due to a study set that included many low-positive samples that were below its limit of detection.


Assuntos
Vírus da Influenza A/isolamento & purificação , Vírus da Influenza B/isolamento & purificação , Influenza Humana/diagnóstico , Técnicas de Diagnóstico Molecular/métodos , Sistemas Automatizados de Assistência Junto ao Leito , Adolescente , Adulto , Idoso , Idoso de 80 Anos ou mais , Criança , Pré-Escolar , Feminino , Humanos , Lactente , Recém-Nascido , Masculino , Pessoa de Meia-Idade , Sensibilidade e Especificidade , Fatores de Tempo , Adulto Jovem