Electrophoresis ; 39(8): 1048-1053, 2018 04.
Artigo em Inglês | MEDLINE | ID: mdl-29384199

RESUMO

Glycosylated hemoglobin (HbA1c) detection is performed routinely in hospitals as it is the most widespread confirmatory diagnosis of diabetes mellitus. Here we present a novel CE method for measuring HbA1c by introducing silica nanoparticles (NPs) modified with a boronic acid derivative (sugar loadings of 51 ± 2 µg/mg) as pseudo-stationary phase. Before the sample injection, SiO2 NP─B(OH)2 were introduced via pressure. Electrophoretic separation was explored through variation of the buffer pH and separation voltage, being the best separation, resolution and shorter separation time achieved with a 25 mM phosphate buffer pH 6.5. The calibration curve obtained was expressed as Area = 182.05%-1 × HbA1c - 377.02; R2  = 0.9826, using a UV/VIS absorbance detector at 415 nm (diode array). No interferences were observed from carbamylated or acetylated hemoglobin and the method shows a noteworthy stability. A paired t-test was applied to compare the developed CE method with a commercial HbA1c test and no significant variations have been observed at a 90% significance level.


Assuntos
Eletroforese das Proteínas Sanguíneas/métodos , Hemoglobinas Glicadas/análise , Eletroforese das Proteínas Sanguíneas/instrumentação , Eletroforese Capilar/instrumentação , Eletroforese Capilar/métodos , Humanos , Nanoestruturas , Reprodutibilidade dos Testes
SELEÇÃO DE REFERÊNCIAS