1.
Int J Latest Trends Eng Technol
; 15(3): 036-41, 2020 Jan.
Artigo
em Inglês
| MEDLINE
| ID: mdl-32219149
RESUMO
A cold atmospheric pressure plasma device was developed using two parallel plates of Low Temperature Co-fired Ceramic with embedded electrodes. The 2.4 cm wide by 1 mm deep plasma discharge operates at 20 kHz with a 2-5 kV AC drive signal across a 0.25 mm gap. Mixed Argon/oxygen plasmas were directed between the plates to flow toward a bacterial biofilm sample for treatment. Results showed that at 4-5 kV the plasma etched away a bacterial biofilm on glass in 10 minutes. In addition, we showed that short plasma treatments rapidly killed biofilm resident bacteria with ED90 values of <15 s.