1.
Radiat Prot Dosimetry ; 193(1): 37-42, 2021 Jan 15.
Artigo em Inglês | MEDLINE | ID: mdl-33723610

RESUMO

Technical performances of two optically stimulated luminescence readers including, a microStar reader and an Auto 200 reader both manufactured by Landauer, Inc. were studied by the National Radiation Protection Agency Dosimetry Laboratory using manufacturer and International Electrotechnical Commission requirements. The two reader tools demonstrated satisfactory results as they met all the requirements. The two readers showed similar performance for the overall uncertainty and the Coefficient of Variation tests. The microStar reader showed better performances for the linearity and quality control tests, whereas the Auto 200 reader has a lower limit of detection. Moreover, an average shift of 0.08 mSv exists when measuring the doses with the two readers.


Assuntos
Luminescência , Radiometria , Doses de Radiação , Incerteza