1.
Rev Sci Instrum ; 82(1): 013906, 2011 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-21280844

RESUMO

A low noise constant current source used for measuring the 1∕f noise in disordered systems in ohmic as well as nonohmic regime is described. The source can supply low noise constant current starting from as low as 1 µA to a few tens of milliampere with a high voltage compliance limit of around 20 V. The constant current source has several stages, which can work in a standalone manner or together to supply the desired value of load current. The noise contributed by the current source is very low in the entire current range. The fabrication of a low noise voltage preamplifier modified for bias dependent noise measurements and based on the existing design available in the MAT04 data sheet is also described.