RESUMO
The present work reports on the development of a new generation of Lab-on-a-chip (LOC) to perform in-situ and real-time potentiometric measurements in flowing water. The device consisted of two differentiated parts: a poly (dimethylsiloxane) (PDMS) microfluidic structure obtained by soft lithography and a fully integrated chemical sensing platform including four working microelectrodes, two reference microelectrodes and one counter microelectrode for detecting ammonium in a continuous mode. The performance of the device was evaluated following its potentiometric response when analyzing ammonium containing samples. As a key parameter, its time of response was compared to that of a commercially available electrical conductivity sensor used as reference sensor during tests in laboratory using flowing tap water and technical scale using flowing wastewater. As a result, the LOC showed a slope of 55â¯mV/decade, a limit of detection of 4·10-5â¯M and a time of full response between 10 and 12â¯s. It was demonstrated that the device can provide fast and reliable data at real time when immersed in a laminar flow of water. Moreover, the test of robustness showed that it was still functional after immersion in sewage for at least 15â¯min. Besides, the LOC reported here can be helpful for a wide variety of flowing-water applications such as aqua culture outlets control, in-situ and continuous analysis of rivers effluents and sea waters monitoring among others.