Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Diabetes Complications ; 29(1): 93-8, 2015.
Artigo em Inglês | MEDLINE | ID: mdl-25264232

RESUMO

BACKGROUND: In bi-hormonal closed-loop systems for treatment of diabetes, glucagon sometimes fails to prevent hypoglycemia. We evaluated glucagon responses during several closed-loop studies to determine factors, such as gain factors, responsible for glucagon success and failure. METHODS: We extracted data from four closed-loop studies, examining blood glucose excursions over the 50min after each glucagon dose and defining hypoglycemic failure as glucose values<60 mg/dl. Secondly, we evaluated hyperglycemic excursions within the same period, where glucose was>180 mg/dl. We evaluated several factors for association with rates of hypoglycemic failure or hyperglycemic excursion. These factors included age, weight, HbA1c, duration of diabetes, gender, automation of glucagon delivery, glucagon dose, proportional and derivative errors (PE and DE), insulin on board (IOB), night vs. day delivery, and point sensor accuracy. RESULTS: We analyzed a total of 251 glucagon deliveries during 59 closed-loop experiments performed on 48 subjects. Glucagon successfully maintained glucose within target (60-180 mg/dl) in 195 (78%) of instances with 40 (16%) hypoglycemic failures and 16 (6%) hyperglycemic excursions. A multivariate logistic regression model identified PE (p<0.001), DE (p<0.001), and IOB (p<0.001) as significant determinants of success in terms of avoiding hypoglycemia. Using a model of glucagon absorption and action, simulations suggested that the success rate for glucagon would be improved by giving an additional 0.8µg/kg. CONCLUSION: We conclude that glucagon fails to prevent hypoglycemia when it is given at a low glucose threshold and when glucose is falling steeply. We also confirm that high IOB significantly increases the risk for glucagon failures. Tuning of glucagon subsystem parameters may help reduce this risk.


Assuntos
Glicemia/efeitos dos fármacos , Diabetes Mellitus Tipo 1/tratamento farmacológico , Glucagon/administração & dosagem , Bombas de Infusão Implantáveis , Adulto , Automação , Peso Corporal , Estudos de Coortes , Diabetes Mellitus Tipo 1/sangue , Diabetes Mellitus Tipo 1/fisiopatologia , Relação Dose-Resposta a Droga , Desenho de Equipamento , Segurança de Equipamentos , Feminino , Glucagon/efeitos adversos , Humanos , Hiperglicemia/prevenção & controle , Hipoglicemia/prevenção & controle , Masculino , Pessoa de Meia-Idade , Estudos Prospectivos , Fatores de Risco , Sensibilidade e Especificidade , Índice de Gravidade de Doença , Resultado do Tratamento
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...