Strahlenther Onkol ; 183(6): 327-31, 2007 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-17520187

RESUMO

BACKGROUND AND PURPOSE: The accuracy of the dose monitor system of a medical electron accelerator is of crucial importance for the precise delivery of the prescribed dose. Special attention is given to the case of low monitor units (MU). The long-term stability of the dose monitor system is analyzed. MATERIAL AND METHODS: Photon and electron fields with monitor units between 1 and 500 MU were measured with field sizes of 10 x 10 cm(2) and 20 x 20 cm(2) at source-skin distances (SSDs) of 100 cm and 95 cm, respectively. The relative dose per MU [see text] was determined for an Elekta linear accelerator. RESULTS: The relative dose per MU was measured to be constant within about 1% down to fields with only 2 MU. The deviation remained below 2% down to 1 MU which is the minimum setting of the accelerator. The long-term stability is better than 0.2% per month. CONCLUSION: No therapeutically relevant deviation of the prescribed dose could be detected for the Elekta linear accelerator. This is a prerequisite for the implementation of intensity-modulated radiotherapy (IMRT) techniques.


Assuntos
Aceleradores de Partículas/instrumentação , Monitoramento de Radiação/instrumentação , Radioterapia Conformacional/instrumentação , Radioterapia de Intensidade Modulada/instrumentação , Carga Corporal (Radioterapia) , Elétrons , Desenho de Equipamento , Humanos , Fótons