Appl Radiat Isot ; 68(4-5): 614-6, 2010.
Artigo em Inglês | MEDLINE | ID: mdl-19815420

RESUMO

This paper presents the measurement of the dose equivalent Hp(10) to low (0.005-1) mSv and high (20-1000) mSv doses by exposure at (241)Am and (173)Cs radiation sources of the halide film with FB-III-D dosimeter system. Accuracy of measurements, standard error of the dose mean value (SEM) and some comments about ability to reread of dosimetric films were determined. A good accuracy was obtained over the important dose ranges. In the low dose range, under 0.1 mSv, the SEM values of the (241)Am doses, recorded on the film under plastic filter, are between -21.36% and +47.51%. For 0.1-500 mSv (137)Cs dose range the SEM values are from -9.55% to +7.24%.