Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Ann Acad Med Singap ; 45(1): 6-11, 2016 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-27118223

RESUMO

INTRODUCTION: This study aimed to determine the reproducibility and usability of anterior segment images taken from a smartphone stabilised on a slit-lamp with those taken from a custom-mounted slit-lamp camera. MATERIALS AND METHODS: This was a prospective, single- blind comparative digital imaging validation study. Digital photographs of patients with cataract were taken using a smartphone camera (an iPhone 5) on a telescopic mount and a Canon EOS 10D anterior segment camera. Images were graded and compared according to the Lens Opacification Classification System III (LOCS III). RESULTS: A total of 440 anterior segment images were graded independently by 2 ophthalmologists, 2 residents and 2 medical students. Intraclass correlation (ICC) between the iPhone and anterior segment camera images were fair for nuclear opalescence (NO) and nuclear colour (NC), and excellent for cortical (C) and posterior subcapsular (PSC) (NO: ICC 0.40, 95% CI, 0.16 to 0.57; NC: ICC 0.47, 95% CI, 0.16 to 0.66; C: ICC 0.76, 95% CI, 0.71 to 0.81; PSC: ICC 0.81, 95% CI, 0.76 to 0.85). There was no difference in grader impression of confidence and images usability between both cameras (P = 0.66 and P = 0.58, respectively). CONCLUSION: Anterior segment images taken from an iPhone have good reproducibility for retro-illuminated images, but fair reproducibility for NO and NC under low light settings. There were no differences in grader confidence and subjective image suitability.


Assuntos
Catarata/diagnóstico por imagem , Fotografação/métodos , Microscopia com Lâmpada de Fenda/métodos , Smartphone , Humanos , Oftalmologia , Estudos Prospectivos , Reprodutibilidade dos Testes , Método Simples-Cego
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...