Rev Mal Respir ; 26(3): 283-90, 2009 Mar.
Artigo em Francês | MEDLINE | ID: mdl-19367202

RESUMO

INTRODUCTION: In the context of reducing endoscopy-related infectious risk and new national guidelines on microbiological samples from bronchoscopy, the results of a surveillance program set up in a hospital center were analyzed. METHODS: Over 4 years, scheduled samples were taken from disinfected bronchoscopes. Bacteriology and mycology tests were used to search for microorganisms. The results were interpreted as falling within three levels: target, alert, and action. Factors that could explain the contamination were studied: age of the bronchoscope, number of uses per year, brand, and model. RESULTS: Out of 96 scheduled samples taken, the compliance rate for the period was 83% and increased (p=0.06) over the 4 years. We identified 15 Pseudomonas (six aeruginosa and nine other species), one Stenotrophomonas, one enterobacterium, and two filamentous fungi. None of the factors studied had a significant effect on sample contamination. CONCLUSION: The microbiological surveillance of bronchoscopes is an indispensable part of the quality assurance of bronchoscope maintenance. It can lead to maintenance of the bronchoscope when a noncompliant result is found.


Assuntos
Broncoscópios/microbiologia , Contaminação de Equipamentos , Humanos , Garantia da Qualidade dos Cuidados de Saúde