Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Talanta ; 249: 123598, 2022 Nov 01.
Artigo em Inglês | MEDLINE | ID: mdl-35688074

RESUMO

In this paper, we present a flexible, inkjet-printed, non-enzymatic lactate sensor with high sensitivity and specificity, using a stacked nickel oxide-Nafion nanocomposite/nickel oxide working electrode. Instead of deploying a pure Nafion film on the top of the nickel oxide particles, the nickel oxide-Nafion nanocomposite layer in the new electrode scheme functions not only as an anti-interfering layer but also a reactive layer and the bottom pure nickel oxide layer free from interfering substances mainly participates in the redox reaction to enhance the sensing current. Experimental results show that the sensor with a working electrode printed using a 30 µL NiOx ink and a mixture of 30 µL NiO and 4 µL Nafion ink can exhibit an anti-interference ability of >95%, a sensitivity of 20.56 nA/mM/mm2, and limit of detection (LoD) of 0.27 mM satisfying the criteria for human lactate detection. In clinical trial, blood plasma test results show that lactate levels detected using this sensor have a strong linear correlation coefficient square of 0.959 with those measured using the colorimetry method used in hospitals, indicating its potential for application in the management of patients with abnormal lactate values requiring intensive care.


Assuntos
Ácido Láctico , Nanocompostos , Eletrodos , Polímeros de Fluorcarboneto , Testes Hematológicos , Humanos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...