We report on 13 cases of periprosthetic stress fracture at the sleeve/stem junction using the Sivash-Range of Motion femoral prosthesis. Radioisotope bone scans confirmed the incidence of fracture, and review of the lateral radiographs revealed anteromedial notching of the distal sleeve on the metaphyseal throat of the femur. Treatment in all cases was expectant with full resolution of symptoms. However, there were 3 cases of recurrence, 1 of which needed revision to a more distally loading stem. This is a rare complication when using this prosthesis, but we recommend a slight alteration of the entry point for the femoral reamer when using this stem and advise nonsurgical management if it occurs, as the natural history is for the condition to settle.


Assuntos
Artroplastia de Quadril/instrumentação , Fraturas de Estresse/epidemiologia , Fraturas de Estresse/etiologia , Prótese de Quadril/efeitos adversos , Fraturas Periprotéticas/epidemiologia , Fraturas Periprotéticas/etiologia , Adulto , Idoso , Artroplastia de Quadril/métodos , Feminino , Necrose da Cabeça do Fêmur/cirurgia , Fraturas de Estresse/diagnóstico por imagem , Luxação Congênita de Quadril/cirurgia , Humanos , Incidência , Masculino , Pessoa de Meia-Idade , Osteoartrite do Quadril/cirurgia , Fraturas Periprotéticas/diagnóstico por imagem , Desenho de Prótese , Falha de Prótese , Radiografia , Estudos Retrospectivos