Interpretation of adsorption kinetics measured with a quartz crystal microbalance (QCM) can be difficult for adlayers undergoing modification of their mechanical properties. We have studied the behavior of the oscillation amplitude, A(0), and the decay time constant, tau, of quartz during adsorption of proteins and cells, by use of a home-made QCM. We are able to measure simultaneously the frequency, f, the dissipation factor, D, the maximum amplitude, A(0), and the transient decay time constant, tau, every 300 ms in liquid, gaseous, or vacuum environments. This analysis enables adsorption and modification of liquid/mass properties to be distinguished. Moreover the surface coverage and the stiffness of the adlayer can be estimated. These improvements promise to increase the appeal of QCM methodology for any applications measuring intimate contact of a dynamic material with a solid surface.


Assuntos
Técnicas Biossensoriais , Proteínas/química , Quartzo , Adsorção , Animais , Técnicas Biossensoriais/instrumentação , Técnicas Biossensoriais/métodos , Adesão Celular/fisiologia , Fenômenos Químicos , Físico-Química , Elasticidade , Eletrodos , Células Epiteliais , Fibronectinas/análise , Cinética , Modelos Químicos , Proteína Estafilocócica A/análise , Células Tumorais Cultivadas , Viscosidade