1.
Clin Chem ; 38(5): 766-8, 1992 May.
Artigo em Inglês | MEDLINE | ID: mdl-1582033

RESUMO

We evaluated a nonisotopic method (CEDIA; cloned enzyme donor immunoassay) for estimating vitamin B12 and folate in serum. The assays were performed with a Cobas-Mira analyzer. Intra-assay CVs were from 3.7% to 11.0% for vitamin B12 and from 1.2% to 10.7% for folate. Interassay CVs ranged from 9.5% to 11.9% for vitamin B12 and from 6.1% to 18.5% for folate. Linearity was satisfactory, with analytical recovery of 94% at 8.7 and 25.4 nmol/L for folate and 280 and 554 pmol/L for vitamin B12. The detection limit was 3.6 nmol/L for folate and 12.3 pmol/L for vitamin B12. Results of this assay correlated well with those of a conventional radioassay: r = 0.98 for vitamin B12 (n = 51) and r = 0.97 for folate (n = 57). The CEDIA was easy to perform but appeared to be unreliable for use with samples from myeloma patients.


Assuntos
Ácido Fólico/sangue , Imunoensaio/métodos , Vitamina B 12/sangue , Humanos , Imunoensaio/normas , Imunoensaio/estatística & dados numéricos , Paraproteínas , Controle de Qualidade